3-9
alarm normally.
4. Install the syringe, set the infusion parameters correctly, and press the Start key
to enter the running screen.
5. The running icon, speed, accumulated volume, and pressure value during the
infusion process are normal.
Battery Powered Test:
1. After the AC power is disconnected for a short time, the battery should be able to
supply power independently.
2. The LED can generate a normal prompt when the battery is not installed or when
the battery is charged or discharged.
WARNING
The battery in this equipment is not removable. Battery replacement
must be performed by a service technician authorized by Shenzhen
Mindray Scientific Co., Ltd.. Installing a fuel cell or replacing the battery
with a person who is not fully trained by Shenzhen Mindray Scientific
Co., Ltd. can result in hazards (Such as overheating, fire, or explosion).
3.6 Electrical Safety Tests
WARNING
The electrical safety test is to verify the electrical safety of the syringe
pump. It is used to detect abnormal electrical hazards. If these hazards
are not detected in time, it may cause harm to the patient or the operator.
All tests can be performed using commercially available safety analyzer
test equipment. Maintenance personnel shall ensure the adaptability,
functional completeness and safety of the test equipment, and be
familiar with their usage.
Electrical safety tests shall comply with the following standards: EN
60601-1 and GB9706.1.
In case of other stipulations in local laws and regulations, implement
electrical safety tests by following relevant stipulations.
All devices driven by AC power and connected to medical instruments in
patient zones must comply with the IEC/EN 60601-1 and GB9706.1
standards. And electrical safety tests on these devices must be
implemented in accordance with the test interval of the monitor.
Содержание BeneFusion SP3
Страница 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Страница 2: ......
Страница 12: ...1 6 FOR YOUR NOTES ...
Страница 24: ...2 12 FOR YOUR NOTES ...
Страница 62: ...5 16 FOR YOUR NOTES ...
Страница 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Страница 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Страница 78: ...7 8 FOR YOUR NOTES ...
Страница 79: ......
Страница 80: ...P N 046 013115 00 3 0 ...