4 - 36
Operator’s Manual
4 Setup
1.
Set the iStorage server properties as described above.
2.
Click [Connect] to verify connection.
3.
Click [Add] to add the service to the Service list.
Edit a network service
Perform the following procedure:
1.
Select the service to be updated in the service list.
2.
You can see properties in the Configure Service area.
3.
Do one of the following to edit the network service:
–
Modify the properties and click [Update] to update the setting.
–
Click [Default] to set as the default service.
–
Click [Delete] to delete it from the service list.
4.8.3 MedTouch/MedSight Preset
You can set environment for MedTouch/MedSight here and then use the MedTouch/MedSight
function by mobile phone or tablet computers. See MedTouch/MedSight manual for details.
4.8.4 Remote Maintenance
NOTE:
•
After the remote maintenance permission is enabled, it indicates that the ultrasound equipment
permits remote access from the remote controller to upgrade the equipment, transmit files, or
check and download logs.
•
Once enabled, the remote maintenance permission is enabled by default after restarting the
ultrasound equipment.
Perform the following procedure:
1.
Select [Remote Maintenance Permission].
2.
Select [Upload logs to remote server] to upload logs to the remote server, Mindray service
engineers can acquire and handle those log.
Check the process in “Task Management” screen.
4.8.5 Q-Path Preset
You can use the ultrasound system to check data on browser directly. After you have ordered
storage service of a network website service, you can check data using the website, authorized
account and password (provided by the service vendor). You can open the browser (Q-View) to
review previously sent DICOM data.
Q-path is a network server provided by Telexy Healthcare Inc. for digital image storage. Q-View is
a client viewing tool for the server. Telexy Healthcare developed technology and a command
structure that allows any Q-view enabled ultrasound system to access Q-path directly from the
ultrasound system using a single control. The primary purpose for Q-view is to provide remote
access to Q-path from the ultrasound system to complete the exam report on the ultrasound system
and submit for QA.
To access Q-Path on the system, the user just opens the Q-View first and then enters the URL, user
account and password provided by Telexy Healthcare, and the system software system will call Q-
View tool then for user Q-Path application.
For details, please contact Q-Path service provider.
Содержание Anesus ME7T
Страница 2: ......
Страница 58: ...This page intentionally left blank ...
Страница 154: ...This page intentionally left blank ...
Страница 164: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 208: ...This page intentionally left blank ...
Страница 254: ...This page intentionally left blank ...
Страница 264: ...This page intentionally left blank ...
Страница 280: ...This page intentionally left blank ...
Страница 311: ......
Страница 312: ...P N 046 018839 00 5 0 ...