GENERAL OPERATION:

The meter is supplied com-
plete with a 9V battery.
Remove the battery com-
partment cover on the back
of the meter and install the
battery while paying atten-
tion to its polarity.

Connect the MA923D/1
probe to the meter.

Turn the instrument on by
pressing the ON/OFF key.
At start-up, the LCD shows
the percentage of the re-
maining battery life for a
few seconds, and then the
current measurement.

Always remove the electrode protective cap
before taking any measurement. If the elec-
trode has been left dry, soak the tip (bottom
2.5 cm) in M10000 rinse solution for a few
minutes to reactivate it.

The temperature reading can be displayed in
°C or °F unit; to select the desired scale,
press and hold the “ON/OFF” key until “TEMP”
and the current temperature unit are dis-
played on the secondary LCD. Use the “SET”
button to select the unit and then press the
“ON/OFF” key a couple of times to return to
normal mode.

To activate the HOLD function, keep pressed
the “HOLD” key. The measured value will be
frozen on the display and the “HOLD” mes-
sage appear on the secondary LCD.

After measurements, switch the meter off
by pressing the “ON/OFF” key. The “OFF”
message appears on the LCD; release the
button.

After use, replace the electrode protective
cap with a few drops of storage solution.

TAKING pH MEASUREMENTS:

Select the “pH” mode with the “SET ” key.

Immerse the tip (2.5 cm) of the probe in the
sample and stir gently.

Read the pH value when the clock sym-
bol stops blinking.

Note: Before taking any  measurement, make

sure that the meter has been calibrated
(the “CAL” tag is displayed on the left
lower corner of the LCD).

pH CALIBRATION PROCEDURE:

It is recommended to recalibrate the meter at
least once every three weeks.
A) Preparation:

Buffer solutions:

1. pH 7.01 / 6.86 (MA9007 / MA9006)
2. pH 4.01 (MA9004) for acidic calibra-

tion (pH < 7) or pH 10.01 / 9.18
(MA9010 / MA9009) for alkaline range
(pH > 7).

Use two beakers for each buffer solu-
tion: one beaker for rinsing the probe,
the other for calibration. In this way the
cross contamination between solutions
is minimized.

B) Procedure:

Select the “pH” mode with the “SET ” key.

Select the calibration buffer set: press
and hold the “ON/OFF” key until the LCD
shows “TEMP”. Press again this key and
the “BUFF” message will appear; then
select the desired buffer set with the
“SET” key: “7.01 pH BUFF” (for standard
solutions: pH 4.01, 7.01, 10.01) or “6.86
pH BUFF” (for NIST solutions: pH 4.01,
6.86, 9.18). Press the “ON/OFF” key again
to exit.

Remove the protective cap, then immerse
the probe in the first buffer solution (7.01
or 6.86 pH).

Press and hold the “ON/OFF” key until
“CAL” is displayed on the lower LCD.

Release the button and the
message “7.01 pH USE”
(or “6.86 pH USE” for NIST
buffer set) will be dis-
played.

USER MANUAL

PORTABLE pH/ORP METER

WITH EXTENDED RANGE

MODEL: SM106

pH, ORP, Temperature Meter