USER MANUAL 

 

pH 55 • pH 56 

Pocket-size 

pH/°C/°F Meters

 

 

 

 

 

www.milwaukeeinst.com 

 
 
 

Automatic Temperature 

Measuring  unit 

Compansation Indicator 

on main LCD 

Instability Indicator 

Main LCD 

 
Calibration Indicator 

Temperature 

Low Battery Indicator 

reading  

 

Secondary LCD 

 
 

OPERATIONAL GUIDE 

• Remove the electrode cap and immerse 

the  meter in 

MA9015 

storage solution for 

2 hours to activate the electrode. 

• Turn the meter on by pressing the 

ON/OFF  button. All the used segments 
on the LCD will be visible for 1 second or 
as long as the button is pressed. 

• Immerse the electrode in the solution to 

be  tested. Stir gently and wait for the 
reading to  stabilize, i.e. the instability 
indicator (hourglass)  on the LCD turns 
off. 

• The pH value is automatically 

compensated for temperature and will be 
displayed on the  main LCD, while the 
temperature is shown  on the secondary 
LCD. 

• 

To freeze the display

, while in 

measurement  mode, press and hold the 
SET/HOLD button. The “HOLD” message 
appears on the secondary  display and 
the reading will be frozen on the LCD. 

Press any button to return to normal mode. 
• 

To turn the meter off

, press the ON/OFF 

button. The “OFF” message will appear 
on  the secondary display. Release the 
button. 

 

Note:  

• Before taking any measurement, make 

sure the meter is calibrated (the CAL tag 
is on). 

• After use always turn the meter OFF, 

rinse the electrode with water to minimize 
contamination  and store it with a few 
drops of storage  (

MA9015

) or pH7 

(

M10007

) solution in the protective cap.  

NEVER USE DISTILLED OR DEIONIZED 

WATER FOR STORAGE PURPOSE. 

 

CALIBRATION PROCEDURE 

For better accuracy, frequent calibration of 

the tester is recommended. Calibration is 
also necessary 

after electrode 

replacement, after testing  aggressive 
chemicals and where extreme accuracy 
is required. 

• From normal operation mode, press and 

hold  the ON/OFF/CAL button until the 
“OFF” message 

on the secondary LCD is substituted by 

“CAL”. Release the button. 

• The instrument enters the calibration 

mode  by displaying “pH 7.01 USE” (or 
“pH 6.86 USE” if the NIST buffer set was 
selected). 

• For a 

single-point calibration

, immerse the 

electrode in any buffer, i.e. pH 4.01, 7.01 
(or 6.86), 10.01 (or 9.18). 

• The meter activates the automatic buffer 

recognition. If no valid buffer is detected, 
the  meter keeps the USE indication 
active for 12 seconds, and then replaces 
it with WRNG,  indicating that the sample 

being measured is  not a valid buffer. 
Otherwise, if a valid buffer  is detected, 
then its value is  shown on the  primary 
display, and REC appears on the 
secondary LCD. 

• If the pH 7.01 (or pH 6.86) was used, 

press  the SET button to exit the 
Calibration mode  and the “OK 1” 
message will appear on the  display. The 
first calibration point is stored  and the 
meter returns to normal measurement 
mode. 

For better accuracy, it is always 
recommended  to perform a 2-point 
calibration. 
 

• 

For a 

two-point calibration

, immerse the 

electrode in pH 7.01 (or pH 6.86) buffer 
solution. 

• 

After the first point has been accepted, 
the meter will then ask for the second 
buffer and the message „pH 4.01 USE” 
will appear. 

• 

Rinse the electrode and immerse it in 
the second solution (pH 4.01, 10.01 or 
9.18). 

• 

If a valid buffer value is detected, the 
REC message is displayed and the 
meter completes the calibration 
procedure. The LCD shows the 
accepted value with the „OK 2” 
message and the instrument returns to 
the normal measurement mode. 
Otherwise, if no valid buffer is detected, 
the meter displays the WRNG 
message. 