USER MANUAL

MW803 • MW804

Pocket-size

pH/EC/TDS/Temperature 

Meters

SPECIFICATIONS
Range

 

 

0.00 to 14.00  pH

 

0 to 3999 μS/cm (MW803)

 

 

0 to 2000 ppm (MW803)

 

0.00 to 20.00 mS/cm (MW804)

 

 

0.00 to 10.00 ppt (MW804)

 

0.0 to 50.0 °C / 32.0 to 122.0 °F

Resolution

 

0.01 pH

 

1 μS/cm / 1 ppm (MW803)

 

0.01 mS/cm / 0.01 ppt (MW804)

 

0.1 °C / 0.1 °F

Accuracy (@25°C)

 

0.05 pH

 

2% FS (EC/TDS) 

 

0.5 °C / ±1 °F

TDS Factor

 

0.45 to 1.00 (CONV)

Temperature Compensation

 

Automatic, with ß=0.0 to 2.4%/°C

Calibration

 

Automatic, 1 point for EC and 1 or 

 

2 points for pH

Probe Replaceable MI60P

Environment

 

0 to 50 °C; 100% RH max.

Battery Type

 

4 x 1.5V; IEC LR44, A76

Battery Life

 

Approx. 100 hours of use

Auto-off

 

After 8 min. of non-use

Dimensions / Weight

 

200 x dia 38 mm / 100 g

Make sure the gasket is in place before scre-

wing  back  the  cap.  Batteries  should  only  be 

replaced in a safe area using the battery type 

specified in this instruction manual.

ACCESSORIES

MI60P

      Replaceable probe for

 

   MW803 & MW804

M10030B

   12.88 mS/cm calibration

 

   solution, 20 ml sachet, 25 pcs

M10031B

   1413 μS/cm calibration solution,          

 

   20 ml sachet, 25 pcs

M10032B

   1382 ppm TDS Calibration Solution, 

 

   20 ml sachet, 25 pcs.

M10038B

    6.44 ppt TDS Calibration, 

 

   20 ml sachet, 25 pcs.

MA9060

     12.88 mS/cm calibration solution,   

 

   230 ml bottle

MA9061

      1413 μS/cm calibration solution,  

 

   230 ml bottle

MA9062

      1382 ppm TDS Calibration 

 

   Solution, 230 mL bottle

M10016B

    Cleaning solution, 20 ml sachet,

 

   25 pcs 

M10000B

    Rinse solution, 20 ml sachet,

 

   25 pcs

M10004B

    pH 4.01 buffer, 20 ml sachet, 

                    25 pcs.

M10007B

    pH 7.01 buffer, 20 ml sachet, 

 

   25 pcs.

M10010B

    pH 10.01 buffer, 20 ml sachet,

 

   25 pcs.

MA9004

      pH 4.01 buffer, 230 ml bottle

MA9007

      pH 7.01 buffer, 230 ml bottle

MA9010

      pH 10.01 buffer, 230 ml bottle

MA9006

      pH 6.86 buffer, 230 ml bottle

MA9009

      pH 9.18 buffer, 230 ml bottle

MA9015

      Electrode storage solution, 230 ml

 

• 

The secondary LCD displays “ESC” for 1 se-

cond and then the meter returns to the normal 

measurement mode.

• 

To  reset  to  the  default  values  and  clear  a 

previous  calibration,  press  the  SET/HOLD 

button  after  entering  the  calibration  mode 

and before the first point is accepted.

• 

The  secondary  LCD  displays  “CLR”  for  1 

second, the meter resets to the default calib-

 

ration and the CAL tag on the LCD turns off.

• 

If a valid buffer value is detected, the REC 

message is displayed and the meter comp-

letes the calibration procedure. 

• 

The LCD shows the accepted value with the 

“OK 2” message and the instrument returns 

to  the  normal  measurement  mode.  Other-

wise, if no valid buffer is detected, the meter 

displays the WRNG message.

 

PROBE REPLACEMENT

• 

Remove the protective cap and unscrew the 

plastic ring on the top of the probe.

• 

Pull out the MI60P probe and replace it with 

a new one.

• 

Make  sure  the  gaskets  are  in  place  before 

screwing back the ring.

BATTERY REPLACEMENT

When the batteries become weak, the battery 

symbol on the LCD will light up to advise that 

only a few hours of working time is remaining.

The meter is also provided with BEPS (Battery 

Error  Prevention  System),  which  avoids  any 

erroneous  readings  due  to  low  battery  level 

by automatically switching the meter off. It is 

recommended to replace the batteries imme-

diately.

To  replace  the  batteries  unscrew  the  battery 

compartment cap and replace all four 1.5V bat-

teries while paying attention to their polarity. 