On Sample Reading 

 

y

“-SA-”, there is too much light 

for the sample measurement. 

Please check if the right sample 

cuvet is inserted.

 

y

“Inv”, the sample and the zero 

cuvet are inverted. 

 

y

“ZErO”, a zero reading was not 

taken. Follow the instruction in 

the measurement procedure for 

zeroing the meter.

 

y

Under range. A blinking “0.00” 

indicates that the sample 

absorbs less light than the zero 

reference. Check the procedure 

and make sure you use the same 

cuvet for reference (zero) and 

measurement.

 

y

A flashing value of the maximum 

concentration indicates an over 

range condition. The concentration 

of the sample is beyond the 

programmed range: dilute the 

sample and re-run the test.

BATTERY REPLACEMENT

Battery replacement must only take place in a non-

hazardous environment.

Simply rotate the battery cover on the back of the 

meter.

Detach the battery from the 

terminals and attach a fresh 

9V battery while paying 

attention to the correct 

polarity. Insert the battery 

and replace the cover.

ACCESSORIES

MI506-100

Free Chlorine (100 tests)

MI0001

Glass cuvets (2 pcs)

MI0002

Caps for cuvets (2 pcs)

MI0003

Stoppers for cuvets (2 pcs)

MI0004

Tissue for wipping cuvets (4 pcs)

MI0005

9V battery (1 pc)

CERTIFICATION

Milwaukee Instruments conform to the CE 

European Directives.

Disposal of Electrical & Electronic 

Equipment.

 Do not treat this product 

as household waste. Hand it over to the 

appropriate collection point for the recycling 

of electrical and electronic equipment.

Disposal of waste batteries.

 This product contains 

batteries. Do not dispose of them with other household 

waste. Hand them over to the appropriate collection 

point for recycling.

Please note: proper product and battery disposal 

prevents potential negative consequences for human 

health and the environment. For detailed information, 

contact your local household waste disposal service 

or go to 

www.milwaukeeinstruments.com

 (USA & 

CAN) or 

www.milwaukeeinst.com

.

RECOMMENDATION

Before using this product, make sure it is entirely 

suitable for your specific application and for the 

environment in which it is used. Any modification 

introduced by the user to the supplied equipment 

may compromise the meter’s performance. For your 

and the meter’s safety do not use or store the meter in 

hazardous environment. To avoid damage or burn, do 

not perform any measurement in microwave ovens.

WARRANTY

These instruments are warranted against defects in 

materials and manufacturing for a period of 2 years 

from the date of purchase. This warranty is limited to 

repair or free of charge replacement if the instrument 

cannot be repaired. Damage due to accidents, misuse, 

tampering or lack of prescribed maintenance is not 

covered by warranty. If service is required, contact 

your local Milwaukee Instruments Technical Service. 

If the repair is not covered by the warranty, you will be 

notified of the charges incurred. When shipping any 

meter, make sure it is properly packaged for complete 

protection. 