Sterylis VS Manual
| 27
The individual parameters are changed using the + and - buttons in the O
3
STERILIZATION section.
The code of the
selected parameter is displayed in the O
3
STERILIZATION section. To move to a given function, select it and press and hold
the PROG.SET button. The selection of a given function will be confirmed by a short beep and all displays will blink.
6.8.1 AUTO-TEST function
The function consists in an automatic check of the operating status of all actuators (fan, UV-C lamps, ozone generators)
and the ozone sensor. If the device detects an error in one of the elements during the process, it will display the error. If an
error is detected during previous use of the unit, the AUTO-TEST function can erase this error if the unit operates correctly
during this function (error erase does not apply to the ozone sensor).
During the active AUTO-TEST function:
The lights blink quickly: safe ozone concentration, warning and operating modes
The display under O
3
STERILIZATION indicates a tESt message
The display under CONCENTRATION indicates the characters - -.- -
The display under OZONING TIME indicates the characters - -.- -
To interrupt the AUTO-TEST function, press any of the START buttons corresponding to filtration, disinfection or
sterilization modes.
When the AUTO-TEST function is completed, the unit will go into standby mode.
6.8.2 OZONE-SENSOR-TEST function
The function consists of an automatic check of the ozone sensor's operating status. If the device detects an error in the
ozone sensor during the process, it will display the error. If an ozone sensor error has been detected during previous use
of the unit, the OZONE-SENSOR-TEST function can erase this error if the unit will function correctly during this function
(applies to the following errors: E500 and E501).
During the OZONE-SENSOR-TEST function: the message O3_t on the display in the O
3
STERILIZATION section, the
measured instantaneous voltage value on the display in the O
3
CONCENTRATION section, the time remaining for checking
the ozone sensor on the display in the OZONING TIME section and the warning lights and the safe ozone concentration
are blinking quickly.
To interrupt the OZONE-SENSOR-TEST function, press any of the START buttons corresponding to filtration, disinfection
or sterilization modes.
When the OZONE-SENSOR-TEST function is completed, the unit will go into standby mode.
6.8.3 PRE-TEST function
The function consists in self-checking the operating status of fans, UV-C fluorescent lamps, ozone generators and
the
ozone sensor. It is always activated when you select and start a particular operating mode. If a negative PRE-TEST result
is obtained, the corresponding error codes are displayed and the device returns to standby mode. In the case of a positive
PRE-TEST result, the device switches to the previously selected operating mode and deletes any previously detected
errors (applies to errors: E101, E102, E103, E500 and E501).
During the active PRE-TEST function:
The green active mode light blinks quickly
Yellow warning pictograms O
3
and ! blink quickly (only when the sterilization mode is selected)
The display under O
3
STERILIZATION indicates the characters "- - -”. / L0CH message in lock mode / selected
AUTO/SET1/SET2/SET3 program in sterilization mode in case of no active errors and blinking tESt message in
case of active errors.