Sterylis VS Manual
| 21
6.3.4 “O
3
sterilisation” mode
CAUTION!
Ozone is a substance hazardous to human and animal health and life. Before starting to operate the
device in sterilization mode, it is essential to read these operating instructions, especially the safety
sections!
NOTE!
Before starting the sterilization mode, replace the activated carbon filter with a standard filter. Such a
filter is included with the device. When the sterilization mode is finished, the filter must be replaced with
an activated carbon filter. For instructions on how to replace the filters, see the section "CHANGING THE
FILTERS".
CAUTION!
In order to carry out the ozoning process correctly and safely, it is necessary to follow the instructions in
Chapter 8 - Procedure for carrying out the ozoning process!
CAUTION!
The sterilization mode can be started after a period of 30 minutes after the device has been switched on. If
you try to start the sterilization before 30 minutes have passed, you will end up with error P005. The error
will be automatically cleared 30 minutes after starting the device and it is possible to start the sterilization
mode. Such protection is related to the time required to heat up the ozone sensor. An ozone sensor that
has not passed the annealing process may indicate the ozone concentration incorrectly. The heating up of
the ozone sensor is indicated by the slow blinking of the safe ozone concentration indicator.
The sterilization mode can be performed in two ways – in automatic mode, where the process is carried out according to
the parameters with factory settings, and in manual mode, where the user can change the parameters of ozoning time
and working ozone concentration.
6.3.4.1 “O
3
sterilisation – automatic” mode
To start O
3
STERILIZATION in automatic mode, press and hold the yellow START button in the O
3
STERILIZATION section
on the user panel. Enter the correct access code.
Entering the access code:
The 4-digit code entered is shown on the display under O
3
STERILIZATION. The + and - buttons change the value by 1
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The
PROG.SET button changes the active digit to the next one on the right. After entering the correct code, confirm it by
pressing the yellow START button longer. If the code is incorrectly entered, the code XXXX or HHHH will be displayed and a
short beep will occur. In this case, the code must be re-entered.
Once the correct code has been approved the display in the O
3
STERILIZATION section will show the selected type of
sterilization, the display in the OZONING TIME section will start counting down the time remaining until the start of
sterilization and the O
3
and ! will start blinking.
After the 120-second countdown, the unit will check the proper functioning of the relevant actuators (PRE-TEST) and
then the ozone generators will start and the unit will operate in the selected mode of operation. If any errors are detected
during the test, the device will display a tESt message, enter the standby mode and then display the corresponding error
codes on the display in the O
3
STERILIZATION section (except for errors F201 and F202, which cause the PRE-TEST to
stop until the error is corrected - the filter cartridge is inserted)
When the device is in sterilization mode, an additional acoustic signal is activated (can be deactivated according to the