Use a TS-3 temperature sensor
only. Do not jumper unused TS-3
terminals.
mA Input
For more information on this feature consult the Transducer (P004) and mA
Input Parameters (P250, P251, and P252) in the EnviroRanger Parameter
Reference.
Rack or Panel Mount
Wall Mount
Level System Synchronization
When multiple ultrasonic level monitors are to be used, ensure the
transducer cable(s) are run in separate grounded metal conduits. Otherwise,
synchronize the level monitors so that no unit can transmit while another is
waiting for echo reception.
To synchronize with another EnviroRanger, or other Milltronics instruments:
•
Mount the level monitors together in one cabinet
•
Use a common power (mains) supply and ground (earth) for all units
•
Interconnect the SYNC terminals of all level monitors.
To synchronize with other Milltronics ultrasonic level monitors:
•
contact Milltronics or your local distributor.
white (-)
black (+)
shield
white (+)
shield
neg.
pos.
shield
mA input
neg.
pos.
shield
mA input