Enhancement Parameters
This section identifies all XPL Plus operator programmable features designed for altering
operation to suit individual operator preferences or overcome measurement difficulties.
Typically, these parameters are only altered as directed by the Troubleshooting Guide . If
RUN mode performance requires improvement, access and alter the following
parameters as required. Otherwise, proceed to Operation.
When an Enhancement Parameter is altered, return to the RUN mode to verify the
performance desired is achieved before making any other changes.
Data Logging Parameters (P300 and P302)
View the maximum temperature recorded in the RUN mode.
Parameter Category
Purpose
Page
Data Logging
to view previous maximum Temperature records
50
Profile Records
intended for use by Siemens Milltronics Service personnel
51
Installation Records
to identify length of service and power failure occurrence
54
Range Calibration
to compensate for measurement offset and/or sound velocity
55
Temperature Compensation
to override automatic temperature compensation features
57
Rate
to override Measurement Response (P003)
59
Measurement Verification
to override Measurement Response (P003)
61
Scanning
to override automatic scanning delay or use Aux. Transducer
63
Display Parameters
to override the preset display operating characteristics
65
SmartLinx Parameters
communications
67
Echo Processing
to troubleshoot false echo detection
68
Advanced Echo Processing
intended for use by Siemens Milltronics Service personnel
70
Test Parameters
intended for use by Siemens Milltronics Service personnel
78
Measurement Parameters
to verify Application Parameter programming
80
Master Reset
to reset numerous Parameter Values
81
P300 (V) TEMPERATURE, TRANSDUCER MAX
Use this feature if vessel temperature is monitored by an Ultrasonic /
Temperature transducer to view the highest temperature recorded (in °C).
values: -50 to 150
Содержание airanger XPL plus
Страница 1: ...Instruction Manual March 2005 XPL PLUS airanger ...
Страница 10: ...Page 6 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Страница 26: ...Page 22 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Страница 82: ...Page 78 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Страница 106: ...Page 102 AiRanger XPL Plus INSTRUCTION MANUAL 7ML19981AN02 ...
Страница 113: ......