Page 12
AiRanger XPL Plus – INSTRUCTION MANUAL
7ML19981AN02
mmmm
m
Inst
allat
ion
Transducer
Temperature Sensor
Use TS-3 temperature sensors only. Don’t jumper the terminals if TS-3’s are not used.
Peripherals
The XPL Plus is compatible with the following Siemens Milltronics peripherals (3 max.):
• SAM-20: 20 relay satellite alarm module (2 max.)
• AO-10: 10 output analog output module (1 only)
• BIC-II: buffered XPL Plus to RS-232C / RS-422 interface (1 only)
Hazardous voltage is present on transducer terminals during
operation.
Transducers
To Transducers (10 max), RG62 A / U Coax, 365 m (1200 ft) / Transducer max.
Transducer cables must be run in a grounded metal conduit separate from
other wiring (except TS-3 temperature sensor wiring, if applicable).
single Temp Sensor
TIB-9 card for multiple Temp Sensors (optional)
XPL Plus
(board B)
to Temp Sensor #1
to additional Temp Sensors
Use Belden 8760 (or equivalent ) 18 AWG, 2 wire shielded / twisted, 365 m (1200 ft) per TS-3 max.