Product Information
16
P
RODUCT
W
ATER
R
ESISTIVITY
S
ENSOR
The Resistivity Sensor is used to measure the electrical resistivity of the post Elix Module Water. The resistivity and the water
temperature are shown on the LCD during PRODUCTION Mode.
If a QGard TL 1 Polishing Pack accessory is added to an Elix System, then the Resistivity Sensor measures the electrical resistivity
after the QGard.
UV
L
AMP
W
HAT DOES THE
UV
L
AMP DO
?
The UV Lamp is powered during PRODUCTION Mode. It emits UV light at 254 nm (Germicidal action) using approximately 10 W
of electrical power. The UV Lamp is used to kill bacteria.
H
OW OFTEN DO
I
CHANGE THE
UV
L
AMP
?
It is recommended to change the UV Lamp every 5000 hours of operation. A timer is incremented every minute the UV Lamp is
powered. The Elix System displays the number of days before the UV Lamp needs replacement. This can be viewed in the DATA
portion of the software.
When the projected lifetime reaches 14 Days, the LCD Message “EXCHANGE UV LAMP IN 14 DAYS” is displayed. This message is
displayed until the UV Lamp is replaced and the software timer is reset.
When the UV lifetime reaches 0 Days, the LCD Message “EXCHANGE UV LAMP OVERDUE: 0 DAYS” is displayed. This message is
displayed until the UV Lamp is replaced and the software timer is reset.
UV
L
AMP
C
ATALOGUE
N
UMBER
See Section 6-1.