![Millipore direct-q 3 R Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/millipore/direct-q-3-r/direct-q-3-r_user-manual_1796915030.webp)
24
Hydrating the System
•
At this time, the Final Filter is not installed. Leave the system
overnight or for several hours in FILLING TANK mode or PRE
OPERATE mode (see
Section Operating Modes
). The system will
regularly recirculate water and rinse off purification media inside
the SmartPak cartridge. Do not leave the system in STANDBY mode.
•
(The next morning) Press the Dispensing Button once to put the
system into DISPENSING mode.
•
Allow about 1 Litre of water to be dispensed from the system.
•
Press the Dispensing Button once again. The system will finish
dispensing water.
•
Remove the Tygon Tubing and the Barbed Fitting from the POU
Dispenser.
•
The Final Filter needs to be installed. See