Decontamination Procedure in an Isolator
1.
Remove the bottle holder from the pump.
2.
Disconnect the drain tubing from the isolator table.
3.
Remove the drain tray from the pump.
4.
Remove the protective cover from the pump head (see
5.
Remove these parts from the isolator.
6.
Disassemble the bottle holder and the drain tray (see
7.
Clean these parts with a nonwoven wipe moistened with one of the
recommended agents listed in this section, then rinse them with clean water.
8.
Autoclave these parts at 121 °C for 30 minutes or 134 °C for 10 minutes.
NOTE
The drain tray and bottle holder should be completely disassembled for
autoclaving.
9.
Return these parts to the isolator. Hang them or place them on a shelf in the
isolator.
NOTE
The drain tray and bottle holder should be completely disassembled during
decontamination in the isolator.
10.
Clean the pump with a nonwoven wipe moistened with one of the
recommended agents listed in this section.
NOTE
It is not necessary to remove the rotor from the pump head unless there has
been an accidental spill.
11.
Start the decontamination procedure.
NOTE
The pump and its accessories are compatible with vaporized hydrogen peroxide
(VHP) and peracetic acid decontamination cycles.
12.
When the decontamination procedure is complete, reassemble the pump (see
).