57
Ready Mode
General information
Purpose
In READY Mode, water can be dispensed from the POD Unit. The System should be left
in READY Mode most of the time.
Display
STANDBY Mode
from READY
Mode
Display
Action
Result
Press
.
READY Mode –
water quality
values
The READY Mode screen display is explained below.
READY Mode screen
Explanation
In this example,
•
The permeate water filling the tank has a
conductivity of 6 µS/cm.
•
The product water dispensed from the POD Unit
has:
−
a resistivity of 18.2 MΩ.cm,
−
is temperature compensated (TC) at 25°C, and
−
the TOC value is 4ppb.
In this example, the System is not dispensing or
recirculating water.