PrestoCHILL
MM138-003
– Operator Manual
69
E.5. Power failure
In the event of a power failure during a chiller program, when the power returns, PrestoCHILL turns on again
and goes back to the home page.
If in manual mode (see chapter 3.6) the chiller program must be restarted by the user.
If in automatic mode (see chapter 3.7) the chiller program restarts after 10 seconds.
In the event of a power failure during a defrosting process in manual mode, when the power returns, the
defrosting process does not restart and the following screen appears:
It is not possible to start with a chiller program: the software informs that it is necessary to run a defrosting
program to clean the working area of the PrestoCHILL.
Press OK and start a defrosting program. Alternatively, dry and thoroughly clean the working area of the
PrestoCHILL and all optional before starting a chiller program.
Before starting, wipe the work plate of the PrestoCHILL with paper or similar absorbent
material to dry it.
In the event of a power failure during a defrosting process in automatic mode, when the power returns, the
defrosting process does not restart and any chiller programs are blocked. The following screen will appear:
Press OK and start a manual defrosting program. Alternatively, dry and thoroughly clean the working area of
the PrestoCHILL and all optional and wait for the PrestoCHILL to run the preset automatic cycle (paragraph
3.7.1).
Before starting, wipe the work plate of the PrestoCHILL with paper or similar absorbent
material to dry it.