Risk of fatal injury from power supply!
– Only use the device in a dry environment.
– With the exception of the service cover, never open the
control unit.
– Never open the power supply.
– Always inform Customer Service (
on page 16) in the event of problems with the device.
– In the event of damage to the insulation, switch off the
power supply immediately and unplug the mains plug.
– Observe the safety rules:
– Pull the mains plug.
– Make sure that nobody reconnects the mains plug to the
power supply.
– Never connect the power supply unit to any voltage other
than 100–230 V ~ 47–63 Hz.
– Only use the genuine power supply unit and the genuine
power supply cable from the manufacturer. Both are avail-
able from Customer Service.
If the power supply unit has to be replaced, only use the
genuine power supply unit (GlobTek Medical/
ITE Power Supply; model no.: GTM96600-6518-T2).
– Only use the genuine rechargeable battery from the manu-
facturer. It is available from Customer Service.
If the rechargeable battery needs to be replaced, use only
the genuine rechargeable battery (Battery Pack –
miha bodytec m.ove; model no.: RRC2054-2).
For information on how to replace the rechargeable battery,
please contact Customer Service (
There is an immediate risk of fatal injury due to electrocution
in the event of contact with live parts. Damage to the insula-
tion or individual components can have fatal consequences.
There is a risk of fatal injury in the event of liquid penetrating
the case!
Risk of fatal injury from improper exposure of the electrodes
to moisture!
– When moistening and putting on the electrodes, ensure
that there is a sufficient distance to the control unit. No
water must penetrate the case.
There is a risk of fatal injury due to electrocution in the event
of liquid penetrating the case of the control unit!
Electrical voltage
L
DANGER
Improper exposure of the elec-
trodes to moisture
L
DANGER
Safety
2021-07-28
miha-bodytec m.ove EMS training device
23