Risk of fatal injury from interactions!
Do not connect athletes/patients to a high-frequency surgical
appliance simultaneously. This can result in burns under the
stimulation current electrodes.
– Keep a minimum distance of 3 ft / 1 m between the device
and a short-wave or microwave therapy appliance. Other-
wise, this can cause fluctuations in the output values of the
stimulation current device.
– Never use the device on people who are fitted with active
(e.g., electronic) medical implants.
Interactions with the device can be dangerous and cause inju-
ries including death. Applying the electrode surfaces in the
vicinity of the chest can increase the risk of ventricular fibrilla-
tion.
So far, there has been no known event of cardiac arrhythmia that
can be attributed to training with the device. Since a theoretical
residual risk remains, strictly observe the instructions below.
Persons affected will lose consciousness after a brief period of
time.
Health risk due to cardiac arrhythmia!
– Never exercise by yourself. Always ensure that a qualified
trainer is paying attention and can intervene in the event of
an emergency.
– Act immediately if the athlete/patient loses consciousness:
– Disconnect the cables from the electrodes. To do so,
unplug the connector plug from the vest.
– Notify emergency medical services.
– Administer CPR until emergency medical services arrive.
– If you have access to a defibrillator, defibrillate the ath-
lete/patient.
The EMS training involves a residual risk which may lead to
cardiac arrhythmia. If untreated, cardiac arrhythmia may result
in death.
Interactions
L
DANGER
Cardiac arrhythmias
L
WARNING
Safety
2021-07-28
miha-bodytec m.ove EMS training device
22