Interval
Component
Cleaning type
As necessary or after use.
Accessories:
n
i-body
®
n
i-body
®
belt
n
i-body
®
straps flex
Disinfection
Chapter 11.2.1 “ Cleaning the i-body
the individual electrodes” on page 104
If visibly dirty, but at least
once a month.
Device
Wipe the control unit, surfaces, and grab
handles with a moist cloth.
Chapter 11.2.3 “Cleaning the control unit”
As necessary, but at least
once a month.
Accessories:
n
i-body
®
n
i-body
®
belt
n
i-body
®
straps flex
Cleaning
Chapter 11.2.1 “ Cleaning the i-body
the individual electrodes” on page 104
Use ISOPROPANOL 70% for disinfection.
The electrode surfaces have an anti-bacterial layer. If necessary,
i-body
®
straps flex, i-body
®
belt, i-body
®
, and additional electrodes
can be cleaned. To do so, proceed as follows:
Personnel:
n
Trainer
1.
Wash the i-body
®
by hand at a maximum water temperature
of 105 °F / 40 °C.
If a washer with a hand wash setting is available, you can
also wash the i-body
®
in the washer using the supplied
laundry net.
2.
Hang the i-body
®
on a clothes hanger and let it dry in a well-
ventilated place.
3.
Wash the i-body
®
belt by hand at a maximum water tempera-
ture of 105 °F / 40 °C.
If a washer with a hand wash setting is available, you can
also wash the i-body
®
belt in the washer using the supplied
laundry net.
4.
Let the i-body
®
belt dry in a well-ventilated place.
5.
Wash the i-body
®
straps flex by hand at a maximum water
temperature of 105 °F / 40 °C.
If a washer with a hand wash setting is available, you can
also wash the i-body
®
straps flex in the washer using the
supplied laundry net.
Observing the cleaning intervals
Cleaning the i-body
®
, i-body
®
straps flex, and the i-body
®
belt
After the training
2021-07-28
miha-bodytec m.ove EMS training device
105