Installation,  

Operation 

& Maintenance

Read and Follow All Safety Instructions. Save These Instructions.

Models MP16A, MP22A, MP36C & MP49C

Extensive Product Information Available at: 

ultraviolet.com

buyultraviolet.com

Owner

s Manual

Owner

s Manual

Содержание MP16A

Страница 1: ...rporation MADE IN THE USA Installation Operation Maintenance Read and Follow All Safety Instructions Save These Instructions Models MP16A MP22A MP36C MP49C Extensive Product Information Available at u...

Страница 2: ...NTY 24 Patent Notice 24 These instructions generally describe the installation operation and maintenance of the Mighty Pure line of water purifiers Models MP16A MP22A MP36C and MP49C Questions that ar...

Страница 3: ...with respect to suitability or performance of this equipment or to the results that may be expected from its use The user should periodically inspect clean as necessary and confirm the presence and g...

Страница 4: ...and flows into the annular space between the quartz sleeve and the chamber wall Suspended microorganisms are exposed to the ultraviolet rays emitted by the germicidal lamp The translucent sight port o...

Страница 5: ...cated in an area where any possible condensation or leakage from the water purifier any purifier accessory and or plumbing will not result in damage to the area surrounding the water purifier For adde...

Страница 6: ...purifier must be connected to the cold water line only Inlet water temperature should not exceed 100 F 5 Installation requires that a 5 micron sediment filter or finer be installed in line prior to th...

Страница 7: ...ber and instead have a PVC pipe installed temporarily for shipping purposes Please follow these instructions if applicable 1 Pull off EASY OFFTM end caps from static gland nut on each end of the water...

Страница 8: ...8 Figure 3 Initial Installation of Quartz Sleeve...

Страница 9: ...conjunction with the Ultraviolet Monitor Safety Sensor or Time Delay Mechanism this valve prevents water flow through the water purifier when an abnormal condition is detected or in the event of powe...

Страница 10: ...purifiers not equipped with the Ultraviolet Monitor lamp operation can be verified by a visible glow through the translucent sight port This provides an indication of lamp operation and does not indi...

Страница 11: ...w exposed 5 Next carefully slide lamp back into chamber until approximately 2 inches of the lamp is exposed on the opposite end Hold lamp and remove lamp socket 6 Lamp should now be disconnected on bo...

Страница 12: ...static gland nut If so remove Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end...

Страница 13: ...er 4 To remove fragments of quartz sleeve hold the purifier vertically and shake The quartz fragments will break and drop out of the purifier through the gland fitting Flush water through chamber bein...

Страница 14: ...nd rinse with clean water Stubborn stains usually can be removed with a dilute hydrochloric acid NOTE Follow all manufacturer s instructions and precautions when handling chemicals Once the quartz rod...

Страница 15: ...ttings Tighten suspect connection or fitting or shut down purifier drain and remove fitting or connection Clean threads reapply thread sealing tape and reinstall Poor purifier performance AND OR Low U...

Страница 16: ...or Optional Optional Optional Optional Flow Control Valve Optional Optional Optional Optional Audio Alarm Optional Optional Optional Optional Solenoid Valve Optional Optional Optional Optional Time De...

Страница 17: ...49C Promate Time Delay Mechanism MP16A through MP49C Quantum Thermal Optimizer MP16A through MP49C Promate Solenoid Valve MP16A 1 2 through MP22A 3 4 MP36C 1 MP49C 1 1 2 MP49C 1 1 2 SureFlo Flow Contr...

Страница 18: ...18 REPLACEMENT PARTS Mighty Pure Models MP16A MP22A Figure 8 Exploded View MP16A MP22A...

Страница 19: ...Quartz Sleeve 1 15 1111A 1 15 1112A 10 Sight Port Plug 1 30 1075 1 30 1075 11 Drain Plug 1 27 1216 1 27 1216 12 Speed Nut 6 32 2 50 1314 2 50 1314 13A Ballast 120v 60Hz 1 10 0137 1 10 0137 13B Ballas...

Страница 20: ...20 Mighty Pure Models MP36C MP49C Figure 9 Exploded View MP36C MP49C...

Страница 21: ...465G4 1 25 1465F5 12 Drain Plug 1 27 1216 1 27 1216 13 Screw 1 4 20 x 3 8 long 4 50 1034 4 50 1034 14 Lock Washer 1 4 4 50 1321 4 50 1321 15 Flat Washer 1 4 4 50 1317 4 50 1317 16 Ballast Housing Cove...

Страница 22: ...22 MAINTENANCE NOTES...

Страница 23: ...23 MAINTENANCE NOTES...