Baskets and inserts
Baskets and inserts should be checked
daily to make sure they are functioning
correctly. The washer-disinfector is
supplied with a check list.
The following need to be inspected:
– Check that rollers on baskets/inserts
are free of hindrance and that they
are secure.
– Check that the basket connector is at
the correct height and screwed on
correctly.
– Check that jets, sleeves and hose
adapters are securely held in
position in baskets/inserts.
– Check that washing solution can flow
unhindered through all jets, sleeves
and hose adapters.
– Make sure caps and closures on
sleeves are correctly located.
– Check that the locking caps in the
module connectors of modular
system mobile units are working
properly.
and where applicable:
– Make sure that the spray arms rotate
freely.
– Make sure the spray arm jets are free
of any blockages. See "Cleaning the
spray arms".
– Does the filter plate in the insert for
hand pieces and contra-angles or in
the E 478 for Sautter cannulae need
changing?
Maintenance