The Wireless Transmitter is intended for use in the electromagnetic environment specified below. The customer
or the user of the Wireless Transmitter should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/-8kV contact
+/-2kV, +/-4kV, +/-8kV,
+/-15kV air
+/-8kV contact
+/-2kV, +/-4kV, +/-
8kV,
+/-15kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/-2 kV for power supply
lines
+/-2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
+/- 1 kV for input/output
lines
Not Applicable
Surge
IEC 61000-4-5
+/-1 kV differential mode
+/-1 kV differential
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
+/-2 kV common mode
+/-2 kV common
mode, 1.25/50 μs
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
For 0.5 cycle
40% U
T
(60% dip in U
T
)
For 5 cycles
70% U
T
(30% dip in U
T
)
For 25 cycles
< 5% U
T
(>95% dip in U
T
)
For 5 s
<5% U
T
(>95% dip in U
T
)
For 0.5 cycle
40% U
T
(60% dip in U
T
)
For 5 cycles
70% U
T
(30% dip in U
T
)
For 25 cycles
< 5% U
T
(>95% dip in U
T
)
For 250 cycles
(5 s @50 Hz)
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Wireless
Transmitter requires
continuous operation during
power mains interruptions, it is
recommended that the
Wireless Transmitter be
powered from an
uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnet fields
should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Table 17.2: Guidance and Manufacturer’s Declaration—ESD and Power Fluctuations
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...