26-0002-0005/0 (en) (2022)
51
• Programming in the presence of an EMI signal (e.g., OR lights, the programming laptop itself,
other computers, or other medical devices). This can be difficult or impossible, but problems
can usually be resolved by repositioning the patient, the WT, or the EMI source.
• Movement of the patient away from the WT during communication
• Proximity of the IPG or WT to a metal table or large metal object
• Presence of metal barriers between the IPG and WT
• IPG battery at end of life (EOL)
• Software (SAPS) needing to be restarted by restarting computer
• Defective WT
• Defective laptop
• Defective IPG
Communication problems are often intermittent and are rarely related to the IPG. The
surrounding environment typically causes these problems.
Restarting the laptop solves most communication and other system issues that are not
connection related.
13.2.2 Unable to Initiate Communication
In certain instances, the user may be unable to communicate with the IPG. After trying all other
remedies (verifying all connections, restarting the computer, etc.), please contact
MicroTransponder for assistance in resetting the IPG. Contact information is provided in
Section 18: Information and Support.
.
14 D
EVICE
L
IFETIME
14.1 IPG Lifetime, Removal, and Replacement
All MicroTransponder IPGs eventually require surgical replacement as a result of battery
depletion. IPG replacement does not, of itself, require Lead replacement unless a Lead
discontinuity (a break) is suspected.
14.1.1 Signs of End of Life for the Vivistim
®
System
The most common reason for the absence of stimulation is battery depletion, although there
may be other reasons. When end of life (EOL) occurs, the IPG will be unable to provide
stimulation. If the IPG is not explanted or replaced at EOL, the battery voltage will continue to
gradually decrease, and communication with the IPG may not be possible.
14.1.2 Replacement Based on Battery Status Indicators
During patient visits, you will use SAPS to check the status of the patient’s IPG (see Figure
4.8: ). The SAPS will display warning messages if the battery is nearing its depletion.
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...