4.6 BIAS
In setting the dc bias voltage, voltage range is ±12VDC, choose↑↑,↑,↓,↓↓,
Adjust it or click the number key (when input is above the instrument voltage
range will be displayed outside range)
4.7 RANGE DC
Select<Range>with<Direction>key. It is preset as<Auto>. The measuring
range belongs to internal parameter, which will be based to search the bay
according to the measuring items. The range recommends setting as<Auto>in
order to obtain better measuring accuracy. The range actually measured will
be displayed at the lower-left corner of LCD panel. Further, there are also <1-
30
Ω all frequency>, <2-300Ω frequency below 1.2 MHz>, <3-3KΩ frequency
below 120 KHz> and<4-30K
Ω frequency below 12 KHz> ranges for option.
Faster measuring speed can be achieved when setting
at “Range Hold”.
4.8 TRIG
Use the <Up/Down> directional buttons to select. The options include <REPEAT> and
<SINGLE>.
4.9 COMP
Use the <Up/Down> directional buttons to select. The options include <Compare
category>.