Chapter 3
Serial Configuration
3-10
MS-850 Raster Scanner User’s Manual
3–Ser
ial
Conf
ig.
Operations Commands
Triggering Mode
Format: <
Kg
triggering mode,filter time
>
Example: To select External Edge, send <Kg3>.
End of Read Cycle
Format: <
Kh
end of read cycle mode,timeout
>
Example: To select Timeout and change the timeout value to 6 seconds, send
<Kh0,600>.
Serial Trigger Character
Format: <
Ki
serial trigger character
>
Example: To define the Serial Trigger Character as a lowercase c, send <Kic>.
External Trigger Level
Format: <
Kj
external trigger level
>
Example: To change External Trigger Level to Negative, send <Kj0>.
triggering mode:
0 = Continuous Read
1 = Continuous Read 1 Output
2 = External Level
3 = External Edge
4 = Serial Data
5 = Serial Data & Edge
trigger filter time (in 128 µs increments):
Default is
78 (9984 µs; approx. 10 ms)
Any number between 2 and 255.
end of read cycle mode:
0 = Timeout
1 = New Trigger
2 = Timeout & New Trigger
timeout (in 10 millisecond increments):
Default is
100 (one second)
.
Any number between 0 and 65535.
serial trigger character:
Default is
^]
.
Any available ASCII character.
external trigger level (initiates a read cycle):
0 = Negative
1 = Positive