Chapter 3
Serial Configuration
3-6
MS-850 Raster Scanner User’s Manual
3–Ser
ial
Conf
ig.
Communications Commands
Host Port
Note
: Changes made in the scanner’s communications parameters such as baud rate,
parity, stop bits, LRC, etc., must be matched in the other device(s) or communications
will be lost. If this occurs, try Autobaud, or do a
Format: <
Ka
baud rate,parity,stop bits,data bits
>
Example: To change Host Port Baud Rate to 2400, send <Ka2>.
Auxiliary Port
Format: <
Ky
mode,baud rate,parity,stop bits,data bits,daisy chain status,
daisy chain ID>
Communications Status Request
Format: <
KT?
>
baud rate:
0 = 600 3 = 4800 6 = 38.4K
1 = 1200
4 = 9600
7 = 57.6K
2 = 2400 5 = 19.2K 8 = 300
parity:
0 = None
1 = Even
2 = Odd
stop bits:
0 = One
1 = Two
data bits:
0 = Seven
1 = Eight
mode:
baud rate:
parity:
stop bits
0 = Disabled
0 = 600
5 = 19.2K
0 = None
0 = One
1 = Transparent
1 = 1200
6 = 38.4K
1 = Even
1 = Two
2 = Half Duplex
2 = 2400
7 = 57.6K
2 = Odd
3 = Full duplex
3 = 4800
7 = 57.6K
4 = Daisy Chain
4 = 9600
8 = 300
data bits:
daisy chain status daisy chain ID
0 = Seven
1 = Eight
0 = Disabled
1 = Enabled
Default is
0
and
/
1 or 2 ASCII characters