Maintenance
Caution
Maintenance may only be performed by qualified
service personnel.
Warning
Risk of corrosive damage!
Batteries contain corrosive electrolyte.
Use requisite protective clothing when working on
batteries.
Warning
Risk of high voltage!
Do not use the battery if the battery monitoring unit
or its cabling are damaged.
Do not touch uninsulated battery terminals,
connections or other live parts.
Contact a service technician.
Warning
Access™ BMU must not be cleaned with a high
pressure washer.
1. Check that the battery monitoring unit’s cables
and current sensor (if available) are free of
faults, in good condition and show no signs of
mechanical damage.
2. Check and rectify any indications on the battery
unit’s control panel.
3. Check that the battery monitoring unit and its
cables are secured firmly on the battery.
4. Check that there is no grime or battery acid on
the Access™ BMU. Clean if necessary.
Recycling
The product is recycled as electronic scrap. Local
regulation apply and should be followed.
Contact information