INTENDED USE

 

Micropoint Biotechnologies Co., Ltd. has optional liquid controls for qLabs

®

 Coag Panel 2 system. 

The qLabs

®

 Coag Panel 2 liquid controls are used with qLabs

®

 Coag Panel 2 Test Strips to verify the 

qLabs

®

 Coag Panel 2 system performance. It is recommended that two levels of liquid controls are 

used to perform the test. 

WARNING AND PRECAUTIONS

 

1. For in vitro diagnostics use only. Do not take internally.   

2. Carefully read instructions before using. Deviations from the described procedure may alter 

performance.   

3. Take all necessary precautions required when handling laboratory reagents.   

4. Do not freeze liquid controls.   

5. Do not use expired liquid controls. 

6. Do not use for any purpose other than descri

bed in the “Intended Use” section. 

 

7. Do not interchange caps as contamination may occur and compromise results.   

8. Refer to local biohazard procedures for safe waste disposal. 

REAGENTS   

Each control contains varied levels of electrolytes, coagulation factors, stabilizers and preservatives. 

The INR and APTT values of qLabs

®

 Liquid Control are lot-specific. 

CONTROL STORAGE AND STABILITY 

 Unopened controls are stable at 2 - 32 ° C until expiration date stated on the label. 

• Once 

opened, controls are stable for 15 days if capped tightly at 2 - 32 ° C. 

• Before sampling, gently swirl the vial several times to ensure homogeneity.

 

Do not freeze. Do not expose to excessive heat. 

QUALITY CONTROL TESTING PROCEDURES 

Read this part together with 

the User’s Manual of qLabs

®  

ElectroMeter, please refer the 

manual if need more detailed meter operation instruction. 

1

.

 

Gather materials

 

Materials provided 

  qLabs

®

 Coag Panel 2 Liquid Control: Level 1 and Level 2 

Materials required (but not provided) 

  qLabs

®

 ElectroMeter 

  qLabs

®

 Coag Panel 2 Test Strips 

  Puncture-resistant container for medical sharps 

2. Insert the strip 

2.1 Turn on the qLabs

®  

ElectroMeter, 

refer to the User’s Manual to enter the QC Mode

, then the 

meter will prompt you to insert a test strip 

2.2 Remove a new test strip from foil pouch. 

2.3 Insert the strip, electrode end first, into the test strip guide. 

3. Enter the Control and Strip information. 

Please follow the User’s Manual of qLabs

®

 ElectroMeter to enter the Control and Strip information 

then continue the next step. 

Always match the Control and Strip information on the display with those on the package. 

 

Control Code                                      Codechip number 

     

4. Wait for the meter to warm up.   

    The meter will warm up automatically for the test. When it is ready, the meter will beep and prompt 

the user to apply control sample. 

5. 

Perform the control test.   

5.1 Before using, slightly swirl the target control at least 10 times and then apply one drop of the 

liquid control to the sample well, the system will start test automatically. 

Do not add more than one drop of liquid control. 

5.2 The QC test result will appear on the screen in approximately 7 minutes (The result interface 

may be different, depending on the software configuration). 

6. Finish the test. 

6.1  Remove  the  used  test  strip  after  the  test  is  completed.  Dispose  of  the  used  test  strip  in 

compliance with the local biohazard regulation. 

6.2 Insert a new test strip and repeat the above procedures to perform the next level liquid control 

test. 

EXPECTED RESULTS

 

The system is working properly and all testing procedure has been done correctly when the test 
results obtained fall within the assigned control INR and APTT range on the vial. 

Please refer to the assigned INR and APTT control range printed on the label of the vial. 

UNEXPECTED RESULTS

 

If the test results fall outside of the assigned range, qLabs

®

 ElectroMeter will display error code       

“E XXX”, check for the following: 

 

1. Controls may be expired and stored improperly. 

2. The user may not be doing the test correctly. Insert a new test strip and repeat the test procedure 

to perform a new test.   

PERFORMANCE CHARACTERISTICS

 

Within-Run   

(N=20) 

INR Value 

  Mean 

S.D. 

%CV 

 

Within-Run   

(N=20) 

APTT Value 

Mean 

S.D. 

%CV 

Level 1 

1.11 

0.02 

2.00 

 

Level 1 

33.03 

1.27 

3.85 

Level 2 

2.64 

0.09 

3.28 

 

Level 2 

64.41 

2.02 

3.14 

 

Between-Run   

(N=40) 

INR Value 

  Mean 

S.D. 

%CV 

 

Between-Run   

(N=20) 

APTT Value 

Mean 

S.D. 

%CV 

Level 1 

1.10 

0.02 

2.02 

 

Level 1 

35.48 

1.19 

3.34 

Level 2 

2.56 

0.10 

3.85 

 

Level 2 

63.80 

3.13 

4.89 

 

ADDITIONAL INFORMATION 

For additional information about this product, please contact the local distributor or Micropoint 

Technical Support by emailing [email protected] or c86 755 21600849. 

 

 

 

 

 

 

 

 

 

 

 

 

P/N 631-62008 Rev. A1 EN 