Page 4
UF AV5500 Area-Velocity Flow Meter
CONNECTIONS
POWER INPUT: 100 to 240 VAC 50/60Hz. No adjustments are necessary for voltages within this
range. Connect L (Live) N (Neutral) and AC Ground.
Optional DC: 9-32 VDC. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: AC power input and relay connection wires must have conduit entry to the
instrument enclosure. Installation requires a switch, overcurrent fuse or circuit breaker in the building (in
close proximity to the equipment) that is marked as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).
FUNCTION TEST
:
Connect the sensor to the sensor terminals as shown on next page, then
apply power. Allow 30 seconds for the UF AV5500 to initialize.
A. Place QZ02L sensor (flat to the bottom) in a bucket of water
about 6” deep and press the
programing button to view
the status page which shows level and velocity. Stir the
water to see a velocity reading.
Содержание UF AV5500
Страница 2: ...Note This page has been left blank intentionally...
Страница 28: ...Page 28 UF AV5500 Area Velocity Flow Meter...
Страница 34: ...Page 34 UF AV5500 Area Velocity Flow Meter Resistance vs Temperature...
Страница 40: ...Page 40 UF AV5500 Area Velocity Flow Meter SS PIPE MOUNTING BAND OPTION VSJ...
Страница 61: ...Page 61 UF AV5500 Area Velocity Flow Meter Dual Non Contacting Sensor Configuration...