Portaflow 222 & 333 User Manual
Micronics
Page 36
Issue 3.1 Feb 20
2. Use the UP/DOWN arrow keys to select
Digital Device 1/2/3 Setup
. Press the ENTER key.
The
Output 1/2/3
menu is displayed.
3. Use the UP/DOWN arrow keys to select
Function
. Press the ENTER key.
4. Use the UP/DOWN arrow keys to scroll through the output types:
Pulse Output, Alarm Output
or Frequency Output
. With the required output selected, press the ENTER key.
5. Edit the settings as required (see Table x).
Pulse Output
Alarm Output
Freq. Output
Setting
Option/default
Setting
Option/default
Setting
Option/default
Quantity Per
Pulse
Volume:0.010 m
3
Energy:36.000kJ
Direction
Rising / Falling
Low Freq.
0 Hz
Pulse
Duration
50ms
Activation
Level
Volume: 500m
3
Flow: 3e+07 l/min
Energy:1.8e+06 kJ
Power: 500kW
Signal: 0.5
Low Value
Flow:0.00 l/min
Power: 0 kW
Signal:0
Contact Type
Normally Open/
Normally Closed
Deactivation
Level
Volume 475m
3
Flow: 2.85e+07 l/min
Energy:1.71e+06 kJ
Power: 475kW
Signal: 0.5
High Freq.
200 Hz
High Value
Flow: 59999996.1 l/min
Power: 1000 kW
Signal: 1
6.2.1
Setting up Volumetric Pulse
The default pulse width is set to 50ms which represents half of one pulse cycle. A 50ms pulse width
is required for most mechanical counters.
Pulse output
switch open
0
50
100
150
200
250
300
350
Time (milliseconds)
Pulse Width
50 milliseconds
Default Pulse Width
Pulse output
switch closed
Содержание Portaflow 222