Real Time Logge
DD-MM-YY
HH:MM:SS
Site name
Site03
File Name
Site03.csv
Logging Interval
5.0 min
Units
mins
Line Ending Format
Unix
Flow Units
l/min
Power Units
MW
->Stop NOW..
Set Auto Start..
Exit
Data Logger
DD-MM-YY
HH:MM:SS
Choose from list of sites..
Site name
QuickStart
Logger Status..
->Copy Log..
Clear log..
List all Logs..
Exit
4.
Duration
shows the logging period calculated from the Start and Stop times.
5. Select
Save Setup & Exit
and press the ENTER key to return to the
Real Time Logger
screen.
5.3
File naming
If you choose to save the existing log it will be saved to the highest number site that does not
currently have an attached log (e.g.
Site19
,
Site18
.... etc). The name of the site to which the log is
saved will be changed to
CopyQS_xx
(where
xx
is a numerical value which is incremented each
time a log is saved). For example: if
Site19
is available when you elect to save the log, the log is
saved to
Site19
and the sitename is changed to
CopyQS_1
. Files are saved in CSV file format.
NOTE: WHEN SETTING UP A NAMED SITE IT IS ADVISABLE TO RUN A BRIEF LOGGING
SESSION ON THE NEW SITE TO ESTABLISH A DATA LOG FILE, AS THIS WILL PREVENT
THE SITE NAME BEING OVERWRITTEN WHEN SAVING THE QUICKSTART LOG, AS
DESCRIBED ABOVE.
5.4
Stopping logging
From the FLOW READING screen, press the Logger function key to access the REAL TIME
LOGGER screen.
1. Press the Logger function key (1) to access the
Real Time Logger
screen.
2. Select STOP NOW to cease logging.
NOTE: THE
STOP NOW
OPTION REPLACES THE
START NOW
COMMAND WHEN LOGGING IS
ACTIVE.
3. Confirm the action when prompted.
4. Select
Exit
to return to the READ FLOW screen.
NOTE: THE LOGGED DATA WILL REMAIN STORED
IN THE INSTRUMENT’S
MEMORY AND
CAN BE ACCESSED AT ANY TIME AS DESCRIBED ABOVE.
5.5
Copying logged data to a USB memory stick
This procedure describes how to copy a stored log
file to a USB memory stick.
1. Connect the memory stick to the Portaflow USB
socket (see page 6).
2. Access the
MAIN
menu.
IMPORTANT: IF THIS IS DONE FROM THE FLOW READING SCREEN ANY LOGGING
CURRENTLY TAKING PLACE WILL BE TERMINATED.
3. Select
Data Logger
from the
MAIN
menu.
4. Select
Choose from list of sites
and select the name of the site to download.
5. When you are ready to begin downloading the log select
Copy log
.
Содержание Portaflow 222
Страница 2: ......
Страница 51: ...Micronics Portaflow 222 333 User Manual Issue 3 1 Feb 20 Page 47...
Страница 56: ...Portaflow 222 333 User Manual Micronics Page 52 Issue 3 1 Feb 20 10 DECLARATION OF CONFORMITY...
Страница 57: ......