Preparing for Analysis
Gemini VII
3-20
239-42828-01 - Dec 2012
7.
Use the sample tube funnel (provided in the accessories kit) and pour the sample from the
weighing container into the sample tube.
8.
Replace the rubber stopper (or seal frit).
9.
Weigh the sample tube set containing the sample and record the value on the Sample Data
Worksheet as
[B] Sample tube set plus sample mass (Before Degas)
.
10. Subtract the
[A] Mass for empty sample tube set
from the
[B] Mass of sample tube set plus sample
and record this value as the
[C] Sample mass (Before Degas)
.
Degassing the Sample
After the sample has been weighed, use a degassing unit to remove any contaminants which may have
adsorbed to the surface or pores. Appropriate degassing units are available from Micromeritics. Refer
to
for ordering information.
If using the SmartPrep degasser, a menu displays on the program menu bar. Degas the sample using
menu commands and information entered in the Degas Conditions window. Refer to the SmartPrep
operator’s manual for operating instructions.
After degassing is complete, perform the following steps:
1.
Weigh the sample tube set containing the sample and record the weight on the Sample Data
Worksheet as
[B] Sample tube set plus sample mass (After Degas)
.
2.
Subtract the
[A] Mass for empty sample tube set (Before Degas)
from the
[B] Sample tube set plus
sample mass (After Degas)
to obtain the sample’s mass. Record this value as
[C] Sample mass
(After Degas)
.
Funnel
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...