Connecting Gases
Gemini VII
9-22
239-42828-01 - Dec 2012
Cleaning and Verifying the Gas Line
Always clean the gas lines and verify there are no leaks at the connections after a gas bottle is con-
nected. This test examines the gas line from the instrument to the gas bottle, then from the instrument
to the regulator shut-off valve. A report is generated at the completion of the test verifying that it has
passed or failed. Causes and corrective action for a failure are provided.
Before beginning, confirm that the state for valves and the low-pressure gauge are as follows:
1.
To start diagnostic testing, go to
Unit [n] > Diagnostics
.
2.
Click the down arrow to the right of the
Test
field and select
Clean and Verify Analysis Gas
Line Rev [n].
This procedure uses the analysis gas as an example. Select
Clean and Verify
Helium Gas Line Rev [n]
if the helium bottle is being verified.
3.
Click
Report after analysis
and select
Preview
as the destination. Click
OK
.
4.
Click
Start
. From the
View
dropdown list, select either
Operation
,
Instrument Log
, or
Instrument Schematic.
.
5.
A series of prompts display on the screen requiring operator response. The length of time a test
will run is also indicated on the screen.
Regulator control knob
OPEN
Gas bottle shut-off valve
CLOSED
Regulator shut-off valve
OPEN
Low-pressure gauge
14 - 15 psig
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...