Calibrating Transducers and Servo Valves
Gemini VII 2390 Confirm
52 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Calibrate the Servo Valves
1. To calibrate the sample servo valve:
a. Pressurize the ports to 780 mmHg as described in step 2 of the previous procedure.
b. Install blank tubes on both ports (refer to
“Installing Sample Tubes” on page 57
c. From the main menu,
Select Unit [n] > Calibration > Sample Servo
.
d. Press
Start
button to begin the calibration routine.
e. When the calibration is complete, a message is displayed. Click
OK.
2. To calibrate the balance servo valve:
a. From the main menu, for the unit number being operated, select
Unit [n] > Calibration
> Balance Servo
.
b. Press
Start
button to begin the calibration routine.
c. When the calibration is complete, a message is displayed. Click
OK.