Installing the Vacuum Pump
Gemini VII 2390 Confirm
18 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Non-Micromeritics Provided Vacuum Pump
Attaching the Hose
Prepare the vacuum pump according to the directions in the vacuum pump operator’s manual.
Follow the instructions for the type of hose you are installing.
• Rubber
• Metal
Rubber Hose
1. Loosen the screws on the Vacuum port on the rear panel of the analyzer. Then remove the
port cap and centering ring.
2. Place the centering ring you removed in Step 1 and the vacuum adapter in the Vacuum
port.
3. Tighten the two port screws by turning them clockwise with a wrench.