Installing the Vacuum Pump
Gemini VII 2390 Confirm
14 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Attaching the Hose
1. Place a centering ring on the oil vapor trap. There are two types of centering rings. Use the
one with the larger opening (as shown below). Then place a hose adapter on the centering
ring.
2. Place the clamp around the vapor trap. Swing the clamp fastener screw around until it fits
into the slot on the other half of the clamp and tighten the wing nut securely by hand. Then
attach the hose to the port.
3. Place a hose clamp on the hose and tighten the retaining nut with a flat-blade screwdriver.
4. Loosen the screws on the Vacuum port on the rear panel, then remove the plug and
centering ring.
Analyzer Rear Panel