Connecting Cables
Gemini VII 2390 Confirm
4 Gemini VII 2390 Confirm Installation Instructions and Checklist
Feb 09
Connecting Multiple Analyzers
An ethernet switch is required when connecting multiple analyzers to the computer.
1. Plug the ethernet switch power cord into an appropriate power source.
2. Connect one end of an ethernet cable to the ethernet port on the computer and the other
end to a numbered port on the ethernet switch.
Do not use the Uplink port.
3. Connect one end of an ethernet cable to the Ethernet connector on the left side of the
analyzer and the other end of the cable to a numbered port on the ethernet switch.
4. Repeat Step 3 for each analyzer you are installing.
Be sure cables are not connected to the
ethernet switch’s uplink port.
Ethernet switch
The finished configuration should look similar to this illustration.