ASAP 2460 Operator’s Manual
Appendix C
246-42800-01 - Aug 2013
C-7
Thermal Transpiration Correction
During data reduction, thermal transpiration correction is applied to the data if the user selected
Apply
thermal transpiration correction
from the
Report Options
window. Starting with the first collected
pressure, the following calculations are performed until the pressure ratio (PC/P) is greater than or
equal to 0.99.
where:
=
Weber’s coefficient, 0.033
=
Weber’s coefficient, 0.245
F
,
Y
,
=
intermediate values for subsequent calculations
G
=
Weber’s coefficient, 2.5
H
=
Weber’s coefficient, 2
MD
=
thermal transpiration hard sphere diameter of gas (Å), from the
Adsorptive
Properties
window
P
=
equilibrated collected pressure measured by gauge at temp
T
amb
SD
=
inside diameter of sample tube (mm), from the
Report Options
window
T
=
average temperature
T
amb
=
room temperature (298 K)
T
bath
=
analysis bath temperature (
K
), from the
Po and Temperature Options
window
Y
P SD MD
2
2.33
T
------------------------------------
10
3
=
1
G
Y
+
1
H
Y
+
-----------------------
=
F
1
Y
2
Y
+
+
---------------------------------
=
P
1
F
1
T
bath
T
amb
------------
–
–
=
T
bath
T
+
2
--------------------
Содержание ASAP 2460
Страница 4: ......
Страница 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Страница 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Страница 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Страница 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Страница 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Страница 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Страница 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Страница 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Страница 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Страница 234: ...Report Examples 2460 Operator s Manual 5 32 246 42800 01 Aug 2013 BJH Adsorption Cumulative Pore Volume...
Страница 235: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 33 BJH Desorption Cumulative Pore Volume...
Страница 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Страница 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Страница 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Страница 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Страница 278: ......
Страница 280: ......
Страница 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Страница 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Страница 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Страница 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Страница 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Страница 415: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 422: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...