8
10.Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating
conditions:
10 - 40 °C / 50 - 104 °F
15-90 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15-90 % relative maximum humidity
Weight:
277 g (including batteries)
Dimensions:
131 x 90 x 60.5 mm
Cuff size:
from 17 - 52 cm according to the cuff
sizes (see «Selecting the correct cuff»)
Measuring
procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V
diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40-199 beats per minute – pulse
Cuff pressure
display range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4 x 1.5 V alkaline batteries;
size AA
Mains adapter DC 6V, 600 mA
(optional)
Battery lifetime:
approx. 920 measurements
(using new batteries)
IP Class:
IP 20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10000 measurements,
whichever comes first
Accessories: 2 years or 5000
measurements, whichever comes first
Содержание BPB2 Basic