6
Clearing all values
If you are sure that you want to permanently remove all stored
values, hold down the M-button (the device must have been
switched off beforehand) until
«CL ALL»
appears and then
release the button. To permanently clear the memory, press the
MAM button while
«CL ALL»
is flashing.
Individual values
cannot be cleared.
Cancel deletion:
press ON/OFF button
1
while
«CL
ALL»
is flashing.
6. Battery indicator and battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AQ
will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol
AQ
will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment
7
at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in Section
«2. Using the device for the first time».
The memory retains all values although date and time must
be reset – the year number therefore flashes automatically
after the batteries are replaced.
Which batteries and which procedure?
Use 4 new, long-life 1.5 V, size AA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
Using rechargeable batteries
You can also operate this device using rechargeable batteries.
Only use «NiMH» type reusable batteries.
Batteries must be removed and recharged when the flat
battery symbol appears. They should not remain inside the
device as they may become damaged (total discharge as a
result of low use of the device, even when switched off).
Always remove the rechargeable batteries if you do not
intend to use the device for a week or more.
Batteries cannot be charged in the blood pressure monitor.
Recharge batteries in an external charger and observe the
information regarding charging, care and durability.
7. Using a mains adapter
You can operate this device using the Microlife mains adapter (DC
6V, 600 mA).
Only use the Microlife mains adapter available as an orig
-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
6
in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
8. Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g.
«Err 3»
, is displayed.
Error
Description Potential cause and remedy
«Err 1»
AS
Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
AR
-B
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
AR
-C
Abnormal
cuff pressure
An adequate pressure cannot be
generated in the cuff. A leak may have
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces
-
sary. Repeat the measurement.
Содержание BP B3 AFIB
Страница 88: ...86 3 2 4 1 5 V AA NiMH 7 Microlife DC 6V 600 mA Microlife 1 6 2 8 Err 3 Err 1 AS Err 2 AR B Err 3 AR C Err 5...
Страница 89: ...87 Microlife BP B3 AFIB GR 9 m H m 12 Err 6 MAM MAM HI 299 mmHg 200 5 LO 40...
Страница 90: ...88 m m...
Страница 99: ...m m 5 HF MRI CT 97 Microlife BP B3 AFIB AR...
Страница 101: ...93 42 EEC IP IP 20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 5 10000 2 5000 99 Microlife BP B3 AFIB AR...
Страница 108: ...9 m m 12 Err 5 Err 6 MAM MAM HI 299 mmHg 200 5 LO 40 106...
Страница 109: ...m m 5 ME HF MRI CT EN60601 1 2 2015 HF 0 3 107 Microlife BP B3 AFIB FA...