Microdyn Nadir MICRODYN BIO-CEL XS-1 Скачать руководство пользователя страница 22

 

 

 

22 | 24 

 

MICRODYN BIO-CEL XS-1 Pilot Plant Manual //  Revision: A // Date: Jul. 25, 2018 

 

 

Cleaning 

Depending on the type of fouling and required cleaning, a sodium hypochlorite solution, citric acid 
solution  or  combinations  of  the  two  may  be  required.  Sodium  hypochlorite  is  used  to  remove 
organic  and  biological  fouling  from  the  membrane  while  citric  acid  is  used  to  remove  mineral 
scaling such as iron, metal salts or calcium salt and other scaling compounds.  
 

 

Important Information 

SODIUM HYPOCHLORITE 

AVOID EXPOSURE TO HEAT AND DIRECT SUNLIGHT! 

STORE IN COOL (T <15°C or 59°F) AND DARK CONDITIONS  

PLEASE NOTE THAT THE ACTIVE CHLORINE CONCENTRATION WILL DEGRADE 
OVER  TIME  AND  RENDER  CLEANING  SOLUTIONS  LESS  EFFECTIVE  OVER 
TIME. 

 

To  safely  perform  chemical  cleanings  and  safely  store  chemicals,  please  follow  the  in-
structions provided by the chemical manufacturer. The pH range and maximum concen-
tration for chemicals used for cleaning purposes should not exceed the limits of the MI-
CRODYN  BIO-CEL

®

  MBR  module  as  listed  on  the  module  data  sheet.  For  further  infor-

mation, please contact the chemical manufacturer. 

In full-scale applications, regular cleaning (maintenance cleaning) is done by dosing chemicals 
directly into the backwash flow of the permeate pipe. The MICRODYN BIO-CEL

®

 XS-1 pilot plant 

has no automatic cleaning integrated, cleaning needs to be done manually.   
 

The frequency of cleaning depends on several factors such as: 

 

Temperature corrected permeability 

 

Type and chemistry of wastewater 

 

Scaling and fouling potential 

 

Membrane conditions (extent of fouling) 

 

Efficiency of production cycles (backwash and relaxation) 

 

Operating temperature 

 

Pretreatment efficiency 

 

Completeness of biological degradation 

 

Permeate flow rates 

 

Aeration intensity 

For  all  municipal  and  most  industrial  applications  with  systems  using  MICRODYN  BIO-CEL

®

 

MBR modules,  the maintenance  cleaning frequencies  are  1-2/week  with  Sodium  Hypochlorite 
and 1/2week with citric acid. 

Please  check  and  note the  normalized  permeability  before  and  after  the  chemical cleaning.  A 
template for a chemical cleaning log book is provided in Chapter 

Fehler! Verweisquelle konnte 

nicht gefunden werden.

 of the MICRODYN BIO-CEL MBR Operation and Maintenance Manual 

and is used to collect essential data for continuous improvement of the cleaning strategy. 

This maintenance cleaning can be done for a biomass/process. The maintenance cleaning pro-
cedure is summarized below: 

  

Table 5

Maintenance Cleaning procedure

*

 

Steps 

Procedure 

 

Stop feed pump, stop filtration. Continue to aerate membranes for 5 minutes. 

Backwash clean water manually. 

i

Содержание MICRODYN BIO-CEL XS-1

Страница 1: ...om www microdyn nadir com en www microdyn nadir com trisep China Office MICRODYN NADIR Xiamen Co Ltd No 66 Jinting North Road Xinglin Xiamen China 361022 infochina microdyn nadir com Singapore Office...

Страница 2: ...N BIO CEL XS 1 10 Membrane 10 Laminate 10 Module 11 4 MICRODYN BIO CEL XS 1 Pilot Plant 12 Equipment 12 Functional Description 13 Connections and Interfaces 14 4 3 1 Waste water sludge supply 14 4 3 2...

Страница 3: ...of the pilot plant 15 Figure 6 Control Panel Overview 19 Figure 7 Control Panel Plant 19 Figure 8 Control Panel System Control 20 Figure 9 Control Panel Parameters 1 20 Figure 10 Control Panel Parame...

Страница 4: ...irectly and indirectly with objects or persons if these defects are attributed to improper handling or to not complying with safety and operating conditions as mentioned in this manual respectively Th...

Страница 5: ...result in death serious injury or equipment damage CAUTION CAUTION indicates a potentially hazardous situation If not avoided may result in injury or equipment damage DANGER PRESSURIZED DEVICE Imprope...

Страница 6: ...mendations for operation and maintenance DO Wear the appropriate personnel protective equipment DO Routinely inspect for wear and tear on equipment and perform maintenance when necessary DO Monitor th...

Страница 7: ...ing and storage are given in Chapter 8 Caution INSTALL THE MEMBRANE MODULE INTO THE TANK DIRECTLY PRIOR TO WATER COMMISSIONING Symbol Instruction Do not place the product under direct sunlight Do not...

Страница 8: ...ring transport assembly and installation Do not separate the membrane sheets if they are attached to each other Do not damage the outer packaging at the top of the product with knives or sharp tools T...

Страница 9: ...of these devices for handling instructions The membrane is preserved and after the preservative is washed out the membrane module should be kept wet at all times to prevent irreversible damage from dr...

Страница 10: ...rane is characterized by an excellent chemical and mechanical durability It is also hydrophilic which allows high flows with a low fouling potential Table 1 NADIR UP150 properties Material Nominal por...

Страница 11: ...tegrated in this module and is located at the base of the module The air is required in order to generate a small amount of cross flow in the module and on the membrane surface in order to be able to...

Страница 12: ...pilot plant Equipment The equipment installed in the pilot plant is listed below The specifications are equal to the specifications in the P ID Figure 5 P ID of the pilot plant Table 3 Overview of eq...

Страница 13: ...filtration tank it is flowing through a 1mm bar screen to remove any coarse materials optional The filtration tank is a round tank with an inner diamerter of 75cm and houses the membrane module It is...

Страница 14: ...e leaser operator will use his own pump or feed supply e g by means of controlling the valve via level differences The feed pipe from the feed source to the connector at the pilot plant is supplied by...

Страница 15: ...igure 3 shows a possible setup for system integration of the BIO CEL XS 1 pilot plant The following works needed to be performed before commissioning date 1 All tanks which are intended to be used for...

Страница 16: ...ion see instruction manual for the plant control a Check the pump performance never run pump with closed V7 and or V9 b Check the blower performance adjust V10 and V11 in a way to achieve 5 6 Nm3 h ai...

Страница 17: ...lowable pressure during backwashing is limited to avoid membrane damage An accumulation of air bubbles in the permeate line due to outgassing can occcur especially during high transmembrane pressures...

Страница 18: ...ification DN Denitrification with intermittent aeration for sludge mixing TDN 0 99 min TDN 30 min Intervall between the aeration times Blower off TDN DELAY 0 99 min TDN DELAY 10 min Aeration for sludg...

Страница 19: ...19 24 MICRODYN BIO CEL XS 1 Pilot Plant Manual Revision A Date Jul 25 2018 Figure 6 Control Panel Overview Figure 7 Control Panel Plant...

Страница 20: ...20 24 MICRODYN BIO CEL XS 1 Pilot Plant Manual Revision A Date Jul 25 2018 Figure 8 Control Panel System Control Figure 9 Control Panel Parameters 1...

Страница 21: ...21 24 MICRODYN BIO CEL XS 1 Pilot Plant Manual Revision A Date Jul 25 2018 Figure 10 Control Panel Parameters 2 Figure 11 Control Panel Chart...

Страница 22: ...irectly into the backwash flow of the permeate pipe The MICRODYN BIO CEL XS 1 pilot plant has no automatic cleaning integrated cleaning needs to be done manually The frequency of cleaning depends on s...

Страница 23: ...ry cleaning is usu ally performed 1 2 times per year and with higher shemnical concentrations The cleaning solu tion must be prepared in an external tank and pumped to the cleaned filtration tank The...

Страница 24: ...he pilot plant 3 Drain the filtration tank and dispose of the sludge through proper channels 4 Rinse and clean all tanks and pipes with permeate water 5 Stabilize the module with a preservative soluti...

Отзывы: