Series-5000 Electric Instruments
Instructions for Use
Страница 1: ...Series 5000 Electric Instruments Instructions for Use ...
Страница 2: ...ymbol Glossary Warnings and Cautions System Components Assembly Disassembly Instructions Reprocessing Instructions Storage Periodic Inspection Repair Service Power Output Noise and Vibration Disposal Warranty 3 4 5 7 7 8 15 15 17 18 18 18 18 18 18 ...
Страница 3: ... show the unlocked status Lock 0018 N A To indicate the function of locking or clamping two machine parts together or location of a machine element in a fixed position Unlock 0019 N A To indicate the function of releasing two machine elements locked or clamped together or the releasing of a machine element from a fixed position Temperature Limitation 0632 ISO 15223 1 20121 Indicates the temperatur...
Страница 4: ...s manufactured The date is expressed as YYYY MM e g 2015 11 or YYYY MM DD e g 2015 11 29 Manufacturer 3082 ISO 15223 1 20121 Indicates the medical device manufac turer 2797 CE Mark for EU Class IIa and Higher Products N A Council Directive 93 42 EEC European Conformity Mark 2797 Notified Body Number Prescription N A FDA Title 21 Chapter 1 Subchapter H Part 801 15 F Federal Law U S A restricts this...
Страница 5: ...nitor the temperature of the nose section WARNING Risk of pinching from moving parts WARNING Risk of wrapped tissue from rotating parts WARNING Risk of kick back from high torque devices WARNING See IM 5025 for operation of the Electric Instrument Control Console WARNING Ensure that there is no electromagnetic interference between the devices in this manual and other devices in use See IM 5025 for...
Страница 6: ...with burs whose head diameter is larger than 4 0mm These burs may break and cause injury WARNING Select the bur guard size for REF 5130 that properly matches the length of the bur WARNING Do not use extra long burs with the REF 5130 CAUTION Remove bur guards from REF 5130 when cleaning WARNING When inserting a blade into REF 5922 the hole in the blade must be seated over the indexing pin to avoid ...
Страница 7: ...World Health Organization recommends processing through a pre vacuum steam sterilization cycle for 18 minutes at 134 C 273 F WHO CDS CSR 2000 3 WHO Infection Control Guidelines for TSE March 1999 WARNING After sterilization allow instruments to cool to room temperature Do not use instruments that are still warm Do not soak instruments or wrap in wet towels to cool SYSTEM COMPONENTS REF Description...
Страница 8: ...able REF 5006 5000 into either of the two console ports marked with an instrument symbol 4 If using a Foot Pedal REF 5401 plug the Foot Pedal Cable REF 5025 5401 into the console port marked with a Foot Pedal symbol Align the marks and gently push the connectors together Only one Foot Pedal can be connected to the console at a time but it can be selected to operate either of the Foot Pedal compati...
Страница 9: ...lace handpiece throttle into the LOCKEDL position when not in use 10 Following the surgical procedure place the handpiece throttle into the LOCKEDL position and disassemble the system for reprocessing 11 Remove single use accessories from the coupler and discard as contaminated sharps 12 Remove the coupler from the handpiece by retracting the collar on the handpiece 13 Remove Electric Cable REF 50...
Страница 10: ...hey may eject and cause injury WARNING Do not use REF 5130 with burs whose head diameter is larger than 4 0mm These burs may break and cause injury WARNING Select the bur guard size for REF 5130 that properly matches the length of the bur WARNING Do not use extra long burs with the REF 5130 CAUTION Remove bur guards from REF 5130 when cleaning Make sure to fully lock the chuck after inserting a bu...
Страница 11: ...e REF 5930 1 If using the REF 5000ET throttle lever handpiece set the throttle safety lock to the LOCKEDL position 2 If using the REF 5000E with a Foot Pedal avoid accidentally activating the Foot Pedal while inserting the bur 3 Select the bur guard size that properly matches the length of the bur Bur Length Bur Guard 47 8mm 66 5mm Medium factory installed and cannot be removed 69 1mm 80mm Long RE...
Страница 12: ...EF 5000E with a Foot Pedal avoid accidentally activating the Foot Pedal while inserting the blade or bur 3 Open the receptacle by turning the chuck counter clockwise 4 Insert the blade or rasp making sure it is fully seated in the locking collar 5 Close and tighten the chuck by turning clockwise 6 To use the instrument set the throttle safety lock to the UNLOCKED position or activate the Foot Peda...
Страница 13: ...rc To insert a blade into the Hall Style Sagittal Saw Module with Key REF 5955 1 If using the REF 5000ET throttle lever handpiece set the throttle safety lock to the LOCKEDL position 2 If using the REF 5000E with a Foot Pedal avoid accidentally activating the Foot Pedal while inserting the blade 3 Use Hex Driver REF 2250 001 to open the blade receptacle by turning the locking nut counter clockwise...
Страница 14: ...trument for 3 5 seconds After it stops set the throttle safety lock to the LOCKEDL position and retighten the locking nut WARNING Make sure that the blade is centered between the washer and the locking plate on REF 5972 and not between just the washer and the locking nut REF 5980 JACOBS DRILL MODULE A low speed high torque rotary device intended for drilling Accepts MicroAire 8051 series and 8054 ...
Страница 15: ... Remove excess body fluids and tissue with a disposable non shedding wipe and cover with a cloth dampened with purified water Body fluids and tissue should not be allowed to dry on instruments prior to cleaning NOTE Instruments and multi use accessories should be cleaned within 30 minutes of use to prevent organic material from drying on the instrument Preparation for Decontamination 1 Remove all ...
Страница 16: ...guards and electric cables thoroughly under running water 50 C 122 F for a minimum of 2 minutes While rinsing use brush REF 5130 BR or similar in lumens and openings 7 Repeat cleaning handpieces couplers bur guards and electric cables thoroughly with warm water 60 C 140 F mild pH enzymatic detergent and a soft brush to scrub all surfaces for 30 seconds paying close attention to crevices 8 Use a po...
Страница 17: ...e World Health Organization recommends processing through a pre vacuum steam sterilization cycle for 18 minutes at 134 C 273 F WHO CDS CSR 2000 3 WHO Infection Control Guidelines for TSE March 1999 WARNING After sterilization allow instruments to cool to room temperature Do not use instruments that are still warm Do not soak instruments or wrap in wet towels to cool ENVIRONMENTAL PARAMETERS OPERAT...
Страница 18: ...f the manufacturer All other expressed or implied warranties are excluded and MicroAire shall have no liability for incidental or consequential damages Repairs or alterations to MicroAire products by anyone other than MicroAire or an authorized MicroAire repair facility will void the product s warranty STORAGE Sterile packaged instruments should be stored in a designated limited access area that i...
Страница 19: ...19 of 20 IM 5000E USA Rev A 2021 01 NOTES ...
Страница 20: ...s Boulevard Charlottesville Virginia 22911 USA In the United States Phone 800 722 0822 Fax 800 648 4309 Outside the United States Phone 434 975 8000 Fax 434 975 4131 www microaire com The following company is not affiliated with MicroAire Surgical Instruments LLC Synthes is a registered trademark of Depuy Synthes USA ...