8
RFT9739 Rack-Mount Transmitter Instruction Manual
Getting Started
continued
To verify the transmitter is in security mode 8:
Use the Scroll button to scroll through process variable screens to event
register screens. If event register screens appear, the transmitter is in
security mode 8. For more information about using the Scroll button and
transmitter display, see Section 6.2, page 60.
To make changes to configuration or calibration parameters once
security mode 8 is entered:
1. Set switches 1, 2, and 3 to the OFF position.
2. Make changes through digital communication or with the Scroll and
Reset buttons (see "Communication configuration mode", page 61).
Custody transfer event registers record changes made to defined
configuration and calibration parameters. For more information about
digital communications, see the following instruction manuals or use
AMS on-line help:
•
Using the HART Communicator with Micro Motion Transmitters
•
Using ProLink Software with Micro Motion Transmitters
•
Using Modbus Protocol with Micro Motion Transmitters
3. Set switches 1, 2, and 3 to the ON position.
To reenter security mode 8:
If security mode 8 has been established previously, and the security
mode has been temporarily changed, it is not necessary to use the
Reset button to reenter security mode 8. In such a case, resetting
switches 1, 2, and 3 to the ON position will reenter security mode 8
immediately.
If a master reset has been performed, it is necessary to use the Reset
button method to reenter security mode 8. See the procedure, above.
To change to a security mode other than mode 8:
1. Perform a master reset (see Section 7.7, page 79, for master reset
procedure).
2. Perform characterization and re-configuration procedures as
described in any of the following instruction manuals or AMS on-line
help:
•
Using the HART Communicator with Micro Motion Transmitters
•
Using ProLink Software with Micro Motion Transmitters
•
Using Modbus Protocol with Micro Motion Transmitters
3. Set switches 1, 2, and 3 to the desired positions. See Table 2-1,
Содержание Fisher Rosemount RFT9739
Страница 1: ...Model RFT9739 Rack Mount Transmitter Instruction Manual Version 3 Transmitters February 2000 ...
Страница 2: ......
Страница 4: ......
Страница 20: ...10 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 26: ...16 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 92: ...82 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 100: ...90 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 110: ...100 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 112: ...102 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 120: ...110 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 122: ...112 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 129: ......