How
to
Set
Up
and
Run
the
FG5
3
When setting the zero position, it is very important to make sure that the
mass is hanging freely, and is not out of range. A substantial variation in
SPHERE OUT voltage when the servo loop is open indicates that the mass
is hanging freely. It is possible for the mass to be out of range of the
detector. In this case, a small positive or negative voltage indicates that
the mass is above or below the detector, respectively. This can happen if
there is a large change in gravity (usually as a result of a large latitude
and/or elevation change) from one site to the next. The zero position can
be set by using the AUTO mode, which will move the mass to the zero
position. One can also move the test mass position using the MANUAL
mode. Before switching to MANUAL mode, first set the trim-pot to move
the motor in the correct direction. The trim-pot should be set above 5 if
the SPHERE voltage is negative or below 5 if the SPHERE voltage is
positive.
52. Once the desired zero-position is reached, (SPHERE output between 0-
20mV) deactivate the motor by turning the front panel knob to OFF. The
spring should again be allowed to settle down for at least two minutes.
Set the coil switch to ON. This switch activates the Superspring main
servo loop for normal operation. At this point, there may still be a
rotation in the test mass( three second period) which is not damped by the
servo. This rotation mode will eventually damp out, and the gravity data
will be become quieter over the first hour.
3.1.14.FRINGE OPTIMIZING
53. To optimize the fringe signal, the test and reference beams must be made
perfectly coincident. The two interfering beams should be perfectly
overlapped and also have no angular deviation for the greatest signal.
The translation of the test beam relative to the reference beam is done by
adjusting the translator plate (sometimes called twiddler). The angular
deviation is minimized by adjusting mirror mount #6 of the
interferometer.
54. Adjust mirror #6, if necessary, until the beams are coincident in the
telescope. Then, move the translator plate (twiddler) until they are
3-17
Содержание FG5
Страница 19: ...Design Components and Function 2 Figure 2 8 Top view of interferometer optics and beam path 2 11 ...
Страница 20: ...Design Components and Function 2 Figure 2 9 Side view of interferometer optics and beam path 2 12 ...
Страница 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Страница 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Страница 32: ......
Страница 42: ...How to Set Up and Run the FG5 3 3 10 ...
Страница 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Страница 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Страница 87: ...Adjustment and Maintenance 4 4 29 ...
Страница 91: ...Adjustment and Maintenance 4 4 33 ...
Страница 104: ...Troubleshooting 5 5 2 ...
Страница 117: ...Troubleshooting 5 5 15 ...
Страница 131: ...Checklists and Logs Appendix D 9 3 ...
Страница 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Страница 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Страница 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...