9.4
Laser Sighting
The thermoMETER CTL operate with a double laser sighting which should support the sensor alignment.
The measurement spot is within the two laser points.
In the focal point of the respective lens
, both laser points overlap each other and therefore mark the
minimum measurement spot as one laser point. This is how the sensor can be positioned for the object to be
measured.
The laser can be activated/ deactivated via the programming keys on the unit or via the software. If the laser
is activated a yellow LED will shine (beside temperature display)
. At ambient temperatures
> 50 °C the laser will switch off automatically.
i
The two laser points mark the position of the measuring spot, but not its exact size. The exact size of
the measurement spot can be found in the optical charts
i
At ambient temperatures > 50 °C the laser will be switched off automatically.
The laser should only be used for sighting and positioning of the sensor, not for permanent use.
> Shorten the lifetime of the sensor diodes
> Inaccurate or incorrect measurements
Laser radiation.
Irritation or injury of the
eyes possible. Close
your eyes or immedi-
ately turn away if the
laser beam hits the eye.
Do not stare into the
laser beam. Avoid
indirect exposure via
reflective surfaces!