Page 49
Set Sensor Parameter
optoNCDT 1420
7.3
Inputs
Go to the menu
Inputs
in the menu
Settings
.
Laser on/off
On / Off
Laser on/off is only effective when pin 8 is connected to
GND.
Multifunctional
input
Zeroing (Mastering) High / Low Sets the function of the digital input. The Trigger influ-
ences capture and export of a measurement value.
Zeroing/Mastering sets the current measurement value
to the entered master value. Teaching scales the analog
output. HTL is defined as active input level.
Trigger In
High / Low
Teaching
Inactive
Key function
Zeroing (Mastering)
Sets the function of the sensor key. Inactive means
keylock.
Teaching
Inactive
7.4
Signal Processing
7.4.1
Preliminary Remark
Go to the menu
Signal processing
in the tab
Settings
.
A diagram appears according to the prior settings in the area
Diagram type
in the right part of the display.
The diagram is active and various settings can be seen immediately. You can find references to the chosen
setting below.
The menus for the area
Signal processing
are located in the left part.
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require you
to specify a value.