Page 107
Appendix| ASCII Communication with Sensor
optoNCDT 1420
A 3.4.4
LASERPOW, Laser Power
LASERPOW FULL|OFF
- FULL: laser power is set to 100 %
- OFF: laser is switched off
1) Available for the sensor models ILD1420-10/25/50.
A 3.4.5
ROI, Video Signal, Masking the Region of Interest (ROI)
ROI <Start> <End>
Set “region of interest”. ROI for start and end is between 0 and 511. “Start” value is smaller than “End” value.
For sensors with 500 mm measuring range this is a individually value < 511.
A 3.4.6
Measurement Value Processing
A 3.4.6.1 AVERAGE, Measurement Value
AVERAGE NONE|(MOVING|RECURSIVE|MEDIAN [<Averaging depth>])
Averaging affects the displacement value to be output.
- MOVING: Moving averaging value (averaging depth 2, 4, 8, 16, 32, 64 and 128 possible)
- RECURSIVE: Recursive averaging value (averaging depth 1 up to 32768 possible)
- MEDIAN: Median (averaging depth 3, 5, 7 and 9 possible)