Page 100

Appendix| ASCII Communication with Sensor

optoNCDT 1420

A 3.3.8 

User Level

A 3.3.8.1  LOGIN, Change of the User Level

LOGIN <Password>

Enter the password to change user level. The following user levels are available:

 - USER (standard user): “read-only” access to all elements and graphical display of output values of web 

surface

 - PROFESSIONAL (expert): “read-only” and “write” access to all elements

A 3.3.8.2  LOGOUT, Change into User Level

LOGOUT

Set user level to USER.

A 3.3.8.3  GETUSERLEVEL, User Level Request

GETUSERLEVEL

Request current user level

A 3.3.8.4  STDUSER, Set Standard User

STDUSER USER|PROFESSIONAL

Set standard user who is automatically logged in after system start. Standard user does not change with 
LOGOUT which means login as standard user is done automatically after the command RESET or power 
supply of sensor is switched on.

A 3.3.8.5  PASSWD, Change Password

PASSWD <Old Password> <New password> <New password>

Change password for user level PROFESSIONAL.
Type in the old password followed by the new password (2x). In case the new password is not typed in cor-
rectly, an error message is will be displayed. Password may only contain letters from A to Z, no numbers 0 to 
9. Watch upper and lower case lettering. The maximum length is limited to 31 characters.

Содержание ILD 1420-10

Страница 1: ...ILD 1420 10 ILD 1420 25 ILD 1420 50 ILD 1420 100 ILD 1420 200 ILD 1420 500 ILD 1420 10 CL1 ILD 1420 25 CL1 ILD 1420 50 CL1 Instruction Manual optoNCDT 1420...

Страница 2: ...MESSTECHNIK GmbH Co KG K nigbacher Stra e 15 94496 Ortenburg Germany Tel 49 0 8542 168 0 Fax 49 0 8542 168 90 e mail info micro epsilon de www micro epsilon com Intelligent laser optical displacement...

Страница 3: ...Data ILD1420 CL1 19 4 Delivery 21 4 1 Unpacking Included in Delivery 21 4 2 Storage 21 5 Installation 22 5 1 Instructions for Installation 22 5 1 1 Reflection Factor of the Target Surface 22 5 1 2 Er...

Страница 4: ...ess via Web Interface 41 6 2 3 Measurement Presentation via Web Browser 43 6 2 4 Video Signal via Web Browser 45 6 3 Programming via ASCII Commands 47 6 4 Timing Measurement Value Flux 47 7 Set Sensor...

Страница 5: ...ng 69 7 5 3 1 Output Scaling 69 7 5 3 2 Output Scaling with Key Select 70 7 5 3 3 Output Scaling via Hardware Input 71 7 5 3 4 Calculation of Measuring Value using Analog Current 72 7 5 3 5 Characteri...

Страница 6: ...ge of the User Level 100 A 3 3 8 2 LOGOUT Change into User Level 100 A 3 3 8 3 GETUSERLEVEL User Level Request 100 A 3 3 8 4 STDUSER Set Standard User 100 A 3 3 8 5 PASSWD Change Password 100 A 3 3 9...

Страница 7: ...Masking the Region of Interest ROI 107 A 3 4 6 Measurement Value Processing 107 A 3 4 6 1 AVERAGE Measurement Value 107 A 3 4 6 2 MASTERMV Mastering Zeroing 108 A 3 5 Data Output 108 A 3 5 1 OUTPUT S...

Страница 8: ...optoNCDT 1420...

Страница 9: ...dware or a software button menu 1 2 Warnings Avoid unnecessary laser radiation to be exposed to the human body Switch off the sensor for cleaning and maintenance Switch off the sensor for system maint...

Страница 10: ...or destruction of the sensor Avoid exposure of sensor to aggressive media detergents cooling emulsions Damage to or destruction of the sensor 1 3 Notes on CE Marking The following apply to the optoNCD...

Страница 11: ...goods are damaged in the event of malfunction or total failure of the controller Take additional precautions for safety and damage prevention in case of safety related applications 1 5 Proper Environm...

Страница 12: ...protection regulations Although the laser output is low looking directly into the laser beam must be avoided Due to the visible light beam eye protection is ensured by the natural blink reflex The ho...

Страница 13: ...s for the manufac turer of laser devices Laser operation is indicated by LED see Chap 5 3 LASER RADIATION DO NOT STARE INTO THE BEAM CLASS 2 LASER PRODUCT IEC 60825 1 2015 07 P 1mW 670nm THIS PRODUCT...

Страница 14: ...R PRODUCT P 0 39 mW 670 nm Laser Klasse 1 nach DIN EN 60825 1 2015 07 Fig 4 Laser warning labels on the sensor cable Fig 5 Laser warning labels on the sensor housing Consequently you can use Class 1 l...

Страница 15: ...h is arranged to the optical axis of the laser beam in a defined angle A signal processor in the sensor calculates the distance of the point of light on the measuring object to the sensor by means of...

Страница 16: ...ptoNCDT 1420 3 2 Auto Target Compensation ATC The Auto Target Compensation ATC enables stable compensation independent of color and brightness of the measuring object Also small objects can be detecte...

Страница 17: ...40 m Measuring rate 3 0 25 kHz 0 5 kHz 1 kHz 2 kHz 4 kHz Light source semiconductor laser 1 mW 670 nm red Permissible ambient light 10 000 lx Spot diameter 10 SMR 90 x 120 m 100 x 140 m 90 x 120 m 750...

Страница 18: ...r LEDs for power and status Key select key for zero teaching factory setting Power supply 11 30 V DC 24 V P 2 W Controller integrated signal processor The specified data apply for a diffusely reflecti...

Страница 19: ...leiterlaser 0 39 mW 670 nm rot Permissible ambient light 10 000 lx Spot diameter 10 SMR 90 x 120 m 100 x 140 m 90 x 120 m MMR 45 x 40 m 120 x 130 m 230 x 240 m EMR 140 x 160 m 390 x 500 m 630 x 820 m...

Страница 20: ...color LEDs for power and status Key select key for zero teaching factory setting Power supply 11 30V DC 24 V P 2 W Controller integrated signal processor The specified data apply for a diffusely refle...

Страница 21: ...ly remove the components of the measuring system from the packaging and ensure that the goods are forwarded in such a way that no damage can occur Check the delivery for completeness and shipping dama...

Страница 22: ...bsequent compensation see Chap 3 2 Dark or shiny objects being measured e g black rubber may require longer exposure times The exposure time is dependent on the measuring rate and can only be increase...

Страница 23: ...t 20 minutes is required to achieve uniform temperature distribution in the sensor If measurement is performed in the micron accuracy range the effect of temperature fluctuations on the sensor holder...

Страница 24: ...ement object In case of traversing measurements surface roughnesses of 5 m and more lead to an apparent distance change Suitable parameters for the averaging number may improve the comparability of op...

Страница 25: ...n 5 only have a disturbing effect with surfaces which are highly reflecting These influences have to be explicitly considered when scanning profiled surfaces Basically the angle be havior of triangula...

Страница 26: ...lled or polished metals that are moved past the sensor the sensor plane must be arranged in the direction of the rolling or grinding marks The same arrangement must be used for color strips Fig 10 Sen...

Страница 27: ...g and operating Mount the sensor by means of 2 screws type M3 or by means of through bores for M2 with the screws from the accessories Bolt connection Through length Screw depth Amount Screw Washer To...

Страница 28: ...mm 30 50 85 150 260 600 Y mm 10 21 28 46 70 190 The indicated free space in the reception area see Fig 13 has to be kept clear from foreign objects and extraneous light of other laser sensors at leas...

Страница 29: ...put yellow RS422 and current output are switched off The RS422 and the current output can be switched on Web interface can also be switched on red Current 4 20 mA measurement value output off Sensor o...

Страница 30: ...PCF1420 x I and PCF1420 x U PCF1420 x I and PCF1420 x U SPS IF2001 USB PCF1420 x I and PCF1420 x U Extension clamp RS422 Fig 15 Connection examples on ILD 1420 The different periphery devices can be c...

Страница 31: ...ls Interface IF2001 USB RS422 USB converter one RS422 IF2004 USB four IF2008 PCI interface card four SPS ILD 1420 or the like Functional input trigger Switch key PLC or the like Switching input laser...

Страница 32: ...rial output Terminate externally with 120 Ohm 6 pink RS422 Tx 7 red U B Supply voltage 11 30 VDC typ 24 VDC P 2 W 8 black Laser on off Switch input Laser is active if input is connected with GND 9 vio...

Страница 33: ...Connection of supply voltage 7 red UB 12 blue Ground 5 4 4 Laser on The measuring laser on the sensor is activated via an HTL switch input This is advantageous if the sensor has to be switched off for...

Страница 34: ...oad shutdown of the output Connect the output 11 white and 12 blue on the sensor to a measuring device Sensor 11 Uout R1 C1 100 nF 12 ILD1420 Iout 7 12 11 30 VDC When using a PCF1420 x U you will get...

Страница 35: ...al wiring for multifunctional input 5 4 7 RS422 Connection with USB Converter IF2001 USB Cross the lines for connections between sensor and PC i Disconnect or connect the D sub connection between RS42...

Страница 36: ...rown UB Push Pull UB 10 brown RL Output is not galvanically isolated 24V logics HTL Imax 100 mA UHmax 30 V saturation voltage at Imax 100 mA Low 2 5 V output GND High 2 5 V output UB Fig 22 Electrical...

Страница 37: ...of the optional accessories see Chap A 1 Mount the 12 pin M12 cable connector if you use a cable carriers suitable sensor cable PCF1420 Avoid excessive pulling to the cables Provide strain relieves n...

Страница 38: ...ched on the sensor runs through an initialization sequence This is indicated by the momentary activation of all the LEDs Once initialization has been completed the sensor transmits a via the RS422 int...

Страница 39: ...22 connection to the sensor exists The sensor is connected to a PC notebook via a RS422 converter supply voltage persists Start the program ILD1420 DAQ Tool Vx x x Fig 24 Auxiliary program for sensor...

Страница 40: ...toNCDT 1420 Choose the desired sensor Click on the button Connect Choose Configuration in the menu Extras Choose the browser type in dialog Configuration and click on the button Fig 25 Dialog configur...

Страница 41: ...ns descriptions of pa rameters and so tips for filling the website Fig 26 First interactive web site after selection of the web interface The sensor is active and supplies measurement values The ongoi...

Страница 42: ...easuring rate averaging and RS422 in blue lettering Changes to the settings are possible by means of the slide Signal quality or by means of the tab Settings The area Diagram type enables the change b...

Страница 43: ...ite measurement distance measurement 1 Stop stops the diagram data selection and zoom function are still possible Pause interrupts recor ding Save opens the Windows selection dialog for file name and...

Страница 44: ...urve points are marked with a circle symbol and the related values are displayed in text boxes above the graphic Peak intensity is also updated 8 The peak intensity is displayed as bar chart 9 Scaling...

Страница 45: ...d 0 distance small on the left and 100 distance big on the right The related measurement value is marked by means of a vertical line peak marking 1 2 3 4 5 6 7 8 9 11 Fig 28 Display of video signals 1...

Страница 46: ...suring and display rate and timestamp are displayed 7 Choice of a diagram type In the setting Auto the diagram type matching the setting is chosen auto matically 8 Mouse over function When moving the...

Страница 47: ...ted either to a serial RS422 interface via a suitable interface converter see Chap A 1 or a plug in card to a PC PLC Pay attention in the programs used to the correct RS422 default setting Once connec...

Страница 48: ...the settings after turning off the sensor 7 2 Overview Parameter The following parameters can be set or changed in the optoNCDT 1420 see tab Settings Inputs Laser on off Multifunctional input Key func...

Страница 49: ...as active input level Trigger In High Low Teaching Inactive Key function Zeroing Mastering Sets the function of the sensor key Inactive means keylock Teaching Inactive 7 4 Signal Processing 7 4 1 Pre...

Страница 50: ...ignal It should prefera bly hit in the area of the highest point peak of the video signal Setting Standard Material with penetration Fig 29 Example Video signals extract with measuring object material...

Страница 51: ...ght and matt measuring ob jects Use a low measuring rate with dark and shiny measuring objects e g black lacquered surfaces to improve the measuring result With a maximum measuring rate of 4 kHz the C...

Страница 52: ...etting no triggering the sensor starts data transmission right after start up Pulse duration of the Trigger in signal must be at least 50 s Triggering of measured value acquisition and output have the...

Страница 53: ...The sensor outputs a fixed number of measurement values when trigger conditions have been met Value range from 1 16383 After termination of data output the analog output sticks to the last value sampl...

Страница 54: ...d immediately before the trigger event cannot be included instead older measurement values are used which were recorded during previous trigger events Activating data recording trigger deactivates dat...

Страница 55: ...the video signal This func tion is used to suppress e g disturbing reflections or extraneous light Evaluation range Measuring range Fig 32 Light blue areas limit the evaluation area The exposure cont...

Страница 56: ...r Handling Error handling adjusts the behavior of the analog output and the RS422 interface in the event of an error Error handling Error output no value Analog output supplies 3 mA instead of the mea...

Страница 57: ...the resolution Eliminate signal spikes or Smooth out the signal Averaging has no effect on linearity In completion of the measuring cycle the internal average is calculated again i The preset average...

Страница 58: ...espectively output value Mode Each new measurement value is added and the first oldest measurement value from the averaging process from the window taken out again This results in short transient reco...

Страница 59: ...low pass behavior Range of values for number of average N is 1 32768 7 4 8 4 Median The median is generated from a pre selected number of measurement values Mode To do so the incoming measurement val...

Страница 60: ...r value is 0 When mastering the sensor s characteristic is parallel displaced The displacement of the characteristic curve reduces the usable measurement range of the sensor the further the master val...

Страница 61: ...stering key select LED state Measuring Green red yellow depends on measuring position yellow Key select Key select t 0 5 min 2 s t 1 t 2 30 ms 3 s The function Zeroing Master ing can be used several t...

Страница 62: ...put can be found in the electrical connec tions see Chap 5 4 6 Fig 35 Flow chart for zeroing mastering hardware input LED state Measuring Green red yellow depends on measuring position yellow Pin 9 vi...

Страница 63: ...You can reduce the measurement output in the sensor if you set the output of every nth measurement value in the web interface or by command Data reductions causes that only n th measurement value is...

Страница 64: ...imum value Value Always 2 points which mark the start and end of a new measuring range are taught Reversal of the output signal is possible with two point scaling Maximum value Value Digital output In...

Страница 65: ...erve d mm MR mm x 1 1 100 102 65520 MR measuring range mm 10 25 50 100 200 500 d distance mm 0 01MR 1 01MR Distance with master ing 18 bits x digital value The output range is also coded with 64235 va...

Страница 66: ...ts x digital value 0 262143 Time stamp 2 words at 16 bit x digital value Lo y digital value Hi 0 65535 0 65535 t ms 65536y x 1 100 t time stamp ms 0 11h55m49 67s Non linearized focus 18 bits x digital...

Страница 67: ...16 measuring range Target at 60 measuring range Target at 60 measuring range SMR EMR 0 16 100 MR Target 10920 8 00 mm SMR EMR 0 60 100 MR Target 39183 30 00 mm SMR EMR 0 60 100 MR Target 39183 30 00...

Страница 68: ...values without zero setting resp mastering Set target at 80 measuring range 40 mm set master value 100 mm 643 0 Reserve Measuring range 50 25 0 Distance after mastering mm 225465 161230 96995 109200 1...

Страница 69: ...ng you can use the digital output see Chap 5 4 8 as a programmable limit switch The measurement object positions for Teach 1 and Teach 2 have to differ from each other The teaching process requires a...

Страница 70: ...t 3 t 4 t 6 t 7 t 8 t 5 yellow yellow min 30 ms min 30 ms Fig 40 Flow chart for output scaling LED state Measuring Green red yellow depends on measuring position Color according to measuring position...

Страница 71: ...Hz flashes green approx 1 Hz Start teaching Position the measuring object to 4 mA Positon the measuring object to 20 mA Teach in 1 Teach in 2 Measuring red t 0 5 min 2 s 2 s 2 s 30 s 30 s t 1 t 3 t 2...

Страница 72: ...20 20 2 EMR reserve d mm MR mm I OUT mA 4 16 MR measuring range mm 10 25 50 100 200 500 d distance mm 0 01MR 1 01MR Current output with mastering reference value is midrange Variables Value range For...

Страница 73: ...R measuring range mm 10 25 50 100 200 500 m n teaching area mm 0 MR d distance mm m n Current output with mastering and teaching Variables Value range Formula I OUT current mA 3 8 4 SMR reserve 4 20 m...

Страница 74: ...100 MR Target 6 56 mA 8 00 mm SMR EMR 0 16 100 MR Target 6 56 mA 8 00 mm SMR EMR 0 60 100 MR Target 13 60 mA 30 00 mm Zero setting master value 0 mm Set master value 5 mm Set master value 10 mm SMR 0...

Страница 75: ...g Out 20 2 mA Standard output Output characteristic after zero setting Output characteristic after m astering 100 Out min Out max Master point Master value Out min Out max 16 8 mm 0 mm 9 44 mA 8 mm 20...

Страница 76: ...put range see Chap 7 5 3 5 SMR EMR 0 60 100 MR Target 13 60 mA 30 00 mm ILD1420 50 Target at 60 set master value 0 mm Set minimum m 20 mm and maximum n 40 mm 0 60 MBA MBE 100 MR Target 12 00 mA 0 00 m...

Страница 77: ...ly in a set of parameters The next time you turn on the sensor they are available again 7 6 2 Unit Language The web interface promotes the units millimeter mm and inch when displaying measuring result...

Страница 78: ...l Expert If an expert logs in the system keylock on the sensor is automatically unlocked Key lock Automatic Range from 1 60 min Value Keylock starts after expiry of defined time Clicking the button Re...

Страница 79: ...ate existing setup Save changes in active setup Define setup after booting Menu New setup Menu Load Save Menu bar Menu Load Save Enter the name for the setup into the field e g rubber 1 21 and click t...

Страница 80: ...he desired setup with the left mouse button area A The dialog Setup management opens Click on the button Export Click on Create setup with the left mouse but ton The dialog Setup management opens Clic...

Страница 81: ...nge a set of parameters by selecting deselecting the check boxes Click on the button Transmit file A Windows dialog for file transmission opens Confirm the dialog with OK The operating system stores t...

Страница 82: ...ependent of the setup and is therefore not together loaded or saved with the setup The following functions are available for the user User Professional Password required no yes Viewing settings signal...

Страница 83: ...reference 7 6 7 Sensor Reset Sensor reset Sensor settings Button Settings for baud rate language unit keylock and echo mode are deleted and the default parameters are loaded Measurement settings Butto...

Страница 84: ...lue is divided into three bytes that differ in the two most significant bits The transmission of additional output values is optional Output value 1 additional L Byte 0 0 D5 D4 D3 D2 D1 D0 M Byte 0 1...

Страница 85: ...tput even while communicating with the sensor For the data transmission with a PC the MICRO EPSILON IF2008 PCI BUS interface card is suitable This can be connected to the sensor via the PCF1420 x IF20...

Страница 86: ...g Therefore an optics anti static brush is suitable or bleeding the screen with dehumidified clean and oil free compressed air Wet Cleaning For cleaning the protective screen use a clean soft lint fre...

Страница 87: ...IF2008 Y adapter cable into an existing or a customized PC software MEDAQLib contains a DLL which can be imported into C C VB Delphi and many additional programs makes data conversion for you works i...

Страница 88: ...al special indirect or incidental damage In the inter est of further development MICRO EPSILON reserves the right to make design changes without notification For translations into other languages the...

Страница 89: ...W rth 691361100006 IF2004 USB 4 channel converter RS422 to USB useable for cable PCF1420 x IF2008 IF2008 Y inclusive driver connections 2 Sub D 1 terminal block PS2020 Power supply for mounting on DI...

Страница 90: ...1420 Extension clamp RS422 EtherCAT extension clamp to connect two ILD1420 sensors with a EtherCAT master Necessary cable PC1402 x CSP IF2008 Y adapter cable Used to connect two sensors with interface...

Страница 91: ...ing Error output no measurement LED State normal operation Key Select Supply voltage red flashes yellow approx 8 Hz flashes yellow approx 1 Hz yellow green t 0 10 15 s t 1 t 3 t 2 10 s t 4 Fig 50 Flow...

Страница 92: ...ex 0A or hex 0D0A The currently set parameter value is returned if a command is activated without parameters The input formats are Command name Parameter1 Parameter2 Command name Parameter1 Parameter2...

Страница 93: ...1 Parameter2 The reply can be used again as command for the parameter setting without changes Optional parameters are only returned if the returning is necessary For example the activated output value...

Страница 94: ...erface Chap A 3 3 7 PRINT Output of all sensor settings User level Chap A 3 3 8 1 LOGIN Change of user level Chap A 3 3 8 2 LOGOUT Change to user in the user level Chap A 3 3 8 3 GETUSERLEVEL User lev...

Страница 95: ...activate digital output Chap A 3 3 10 7 ERRORHYSTERESIS Stable digital output Chap A 3 3 10 8 ERROROUTHOLD Minimum duration active digital output Handling of setups Chap A 3 3 11 1 IMPORT Load paramet...

Страница 96: ...rocessing Chap A 3 4 6 1 AVERAGE Selection of measurement averaging Chap A 3 4 6 2 MASTERMV Mastering Zeroing Data Output General Chap A 3 5 1 OUTPUT Selection measuring value output Chap A 3 5 2 OUTR...

Страница 97: ...ILD1420 10 Model name sensor sensor series Serial 15030002 Serial number Option 000 Option number of sensor Article 4120212 Article number of sensor Cable head Wire Measuring range 10 00mm Measuring r...

Страница 98: ...r RESETCNT TIMESTAMP MEASCNT Sets back internal counters in the sensor TIMESTAMP sets back timestamp MEASCNT sets back measurement counter A 3 3 6 ECHO Switching the Command Reply ASCII Interface ECHO...

Страница 99: ...OUTREDUCEDEVICE NONE BAUDRATE 921600 OUTREDUCECOUNT 2 UNIT MM OUTVIDEO_RS422 NONE LANGUAGE DE OUTADD_RS422 NONE MFIFUNC NONE GETOUTINFO_RS422 DIST1 MFILEVEL HTL_HIGH OUTHOLD NONE KEYFUNC NONE ERROROUT...

Страница 100: ...nt user level A 3 3 8 4 STDUSER Set Standard User STDUSER USER PROFESSIONAL Set standard user who is automatically logged in after system start Standard user does not change with LOGOUT which means lo...

Страница 101: ...nts OUTPUT triggers the measurement value output Measurement values immediately before the trigger event are included when calculating the mean A 3 3 9 3 MFILEVEL Input Pulse Multifunctional Input MFI...

Страница 102: ...interface MM representation in mm INCH representation in customs A 3 3 10 3 MFIFUNC Function Selection Multifunctional Input MFIFUNC NONE MASTER TEACH TRIGGER Choose function of the multifunctional i...

Страница 103: ...or Wiring of digital output ERROR1 see Chap 5 4 8 A 3 3 10 6 ERRORLIMIT ERRORLIMIT DIST1 upper threshold Digital output is activated if the defined value for a measurement is exceeded Range 0 2 measur...

Страница 104: ...a prompt occurs A 3 3 11 3 MEASSETTINGS Load Save Measurement Settings MEASSETTINGS Subcommands Name Settings of the measurement task Loads proprietary presets and user specific settings from the sen...

Страница 105: ...standard preset for the measurement set ting resp the the default parameter for the device settings are being loaded MEASSETTINGS Measurement settings are being deleted and the standard preset is bei...

Страница 106: ...oot A 3 4 Measurement A 3 4 1 TARGETMODE Measurement Mode TARGETMODE STANDARD MULTISURFACE PENETRATION 1 Choice of material dependent presets see Chap 7 4 2 STANDARD suitable for materials e g made of...

Страница 107: ...rt and end is between 0 and 511 Start value is smaller than End value For sensors with 500 mm measuring range this is a individually value 511 A 3 4 6 Measurement Value Processing A 3 4 6 1 AVERAGE Me...

Страница 108: ...asurement value a maximum of 2 seconds and masters it If no measurement value is received within this time for example by external triggering the command returns with the error E104 Timeout The master...

Страница 109: ...f measurement cycles then an error value is output n 1 1024 A 3 5 5 Selection of Measurement Values to be Output A 3 5 5 1 GETOUTINFO_RS422 Request Data Selection GETOUTINFO_RS422 The command lists al...

Страница 110: ...ent measurement MEASRATE Measuring rate under consideration of reflectivity and movement of the target AVERAGE Averaging of the measurement value under consideration of reflectivity structure and move...

Страница 111: ...E232 Wrong parameter count Too high or too small number of parameters E234 Wrong or unknown parameter type A transmitted parameter has a wrong type or a wrong num ber of parameters were transmitted E...

Страница 112: ...r the measurement setting begins or ends with spaces or is empty E362 Storage region is full Number of storable measurement settings is reached E363 Setting name not found Name of the measurement sett...

Страница 113: ...anced dynamic The signal quality affects averaging of measurement values A 4 2 Tab Settings A 4 2 1 Inputs Laser on off On Off Laser on off is only effective when pin 8 is connected to GND Multifuncti...

Страница 114: ...vel Continuous measurement output takes place as long as the selected level is applied Level selection see Chap 7 3 Pulse duration must at least be one cycle time The subsequent break must also be at...

Страница 115: ...the last valid value Hold last value 1 1024 Value Measurement value averaging No averaging Measurement values are not averaged Moving N values 2 4 8 128 Value Indication of averaging mode The averagin...

Страница 116: ...sible End of range Value Digital output Inactive Regulates the switching performance of the digital output Error see Chap 5 4 8 Analog range digital output switches when the scaled analog range is exc...

Страница 117: ...Page 117 Appendix Control Menu optoNCDT 1420 LED Output Web interface RS422 Current output Parametrization Measurement chart Selected output interface Web inteface yellow RS422 green Analog red...

Страница 118: ...nt set tings Create setup Setup 1 Setup 8 Load Activates a saved measurement setting setup Save Saves changed measurement settings to an existing setup Favorite Selects a setup which is used after reb...

Страница 119: ...ng of existing settings Apply settings of the imported boot setup Transmit data Access permission Current access permission Value Read only Logout Login Button starts change of access permission User...

Страница 120: ...aded Reset all By clicking the button the settings for the sensor measurement settings access permission password and setups are deleted The 1st preset is loaded Restart sensor By clicking the button...

Страница 121: ......

Страница 122: ...N MESSTECHNIK GmbH Co KG K nigbacher Str 15 94496 Ortenburg Germany Tel 49 0 8542 168 0 Fax 49 0 8542 168 90 info micro epsilon de www micro epsilon com X9751351 B021088MSC X9751351 B02 MICRO EPSILON...

Отзывы: