Page 61
Set Sensor Parameter
optoNCDT 1320
7.5.3.3
Output Scaling via Hardware Input
Scaling of the analog output can be made via an impulse at the functional input, the violet wire on the sensor
cable.
violet wire
LED
state
Error
Measuring
Green, red, yellow,
depends on
measuring position
Color according
to measuring
position
flashes red
approx. 1 Hz
flashes green
approx. 1 Hz
Start
teaching
Position the
measuring object
to 4 mA
Positon the
measuring
object to 20 mA
Teach-in 1
Teach-in 2
Measuring
red
t 0
5 min
2 s
2 s
2 s
30 s
30 s
t 1
t 3
t 2
t 4 t 5
t 7 t 8
t 9
t 6
yellow
yellow
min.
30 ms
min.
30 ms
min.
1 ms
Fig. 37 Flow chart for output scaling
LED
state
Measuring
Green, red, yellow,
depends on
measuring position
Color according
to measuring
position
Flashes red approx. 1 Hz
Measuring
red
t 0
5 min
2 s
200 ms
t 5 - t 3 = 2 s
5 ... <10 s
t 1
t 2
t 3 t 4
t 5
yellow
Error
violet wire
Fig. 38 Flow chart for the return of output scaling