Page 101
Warranty
optoNCDT 1320
A 4.2.2
Signal Processing
Measurement task
Standard
Suitable for materials made of ceramics, metal or filled plastics
Changing surface
1
Suitable for e.g. Printed circuit boards (PCB) or hybrid materials
Material with penetration
1
Suitable for plastics (POM, Teflon), materials with strong penetration of
the laser
Measuring rates
250 Hz / 500 Hz / 1 kHz /
2 kHz
For light and matt measuring objects a high measuring rate may be
required. For dark or bright measuring objects a slower measuring rate
may be required (e.g. black lacquered surface) to improve the measu-
ring result.
Output trigger
Level
Continuous measurement output takes place as long as the selected
level is applied. Level selection, see Chap.
. Pulse duration must at
least be one cycle time. The subsequent break must also be at least one
cycle time.
Edge
infinite
Edge selection. „0“ end trigger, „1 ... 16382“ values per trigger, „16383“
infinite trigger
manual
Number
Value
Software infinite
Software triggering is started by clicking the button
Release trigger
.
„0“ end trigger, „1 ... 16382“ values per trigger, „16383“ infinite trigger
manual
Number
Value
Inactive
No triggering
1) Available for the sensor models ILD1320-10/25/50.